DMTA041-01EN [U8778479], Rev. C, May 2014
Chapter 1
16
Figure 1-1 The TomoScan FOCUS LT front panel
The front panel of the TomoScan FOCUS LT contains the following elements:
I/O
This connector is used to connect digital input/output signals.
DC OUT
This connector provides a power supply voltage of ±12 VDC rated at 1 A for
external devices.
DC IN
This connector allows the TomoScan FOCUS LT to be supplied from a DC power
source. The
DC IN
connector accepts voltages in the range of 15 VDC to 24 VDC,
rated at 5 A.
The external DC power supply must be able to withstand up to 10 A of “in rush
current” for a few microseconds.
UT ON
This button alternately enables and disables the
P
connectors.
When this indicator light is on, it indicates that the conventional ultrasound mode
is enabled and that the last four phased array channels are routed onto the
P125
,
P126
,
P127
, and
P128
connectors.