DMTA041-01EN [U8778479], Rev. C, May 2014
Operation Overview
33
3.4
Ethernet Link
The TomoScan FOCUS LT must be linked to the workstation through an Ethernet
network. To make this possible, the TomoScan FOCUS LT is equipped with internal
Ethernet hardware. For a network configuration using a network hub, a standard
Ethernet cable can be used. For a point-to-point configuration, an Ethernet crossover
(category 5e or higher) cable is used.
The TomoScan FOCUS LT is designed to be used with a 100BASE-T Fast Ethernet
network.
The TomoScan FOCUS LT has no internal disk drive. The instrument must therefore
be linked, through an Ethernet network, to a workstation running a Bootp server
software application before it can be turned on, so that all the necessary data for
proper functioning can be loaded. Once booted, the internal RPC (remote procedure
call) server of the TomoScan FOCUS LT connects to the workstation (that is, the client)
through the Ethernet network. For more details on installing and using the acquisition
software relating to the TomoScan FOCUS LT, refer to the
TomoView Reference Manual
.
Summary of Contents for TomoScan FOCUS LT
Page 6: ...DMTA041 01EN U8778479 Rev C May 2014 Table of Contents vi...
Page 10: ...DMTA041 01EN U8778479 Rev C May 2014 Labels and Symbols 4...
Page 18: ...DMTA041 01EN U8778479 Rev C May 2014 Important Information Please Read Before Use 12...
Page 30: ...DMTA041 01EN U8778479 Rev C May 2014 Chapter 1 24...
Page 40: ...DMTA041 01EN U8778479 Rev C May 2014 Chapter 3 34...
Page 64: ...DMTA041 01EN U8778479 Rev C May 2014 Chapter 5 58...
Page 74: ...DMTA041 01EN U8778479 Rev C May 2014 List of Figures 68...
Page 76: ...DMTA041 01EN U8778479 Rev C May 2014 List of Tables 70...
Page 82: ...DMTA041 01EN U8778479 Rev C May 2014 Index 76...