Fig. 3

Adjusting  the  Parfocality  of  the  Observation  Image  and  Monitor 
Image (Fig. 2)

} The parfocality adjustment makes it unnecessary to correct focusing when 

the observation image is switched to the monitor image.

  Use the Allen screwdriver provided with the microscope for this adjust-

ment.

  The parfocality adjustment range is ±1.5 mm. If the adjustment 

on this adapter is not enough, please also use the focusing 

function of the camera.

1. Using  a  high-power  objective,  look  into  the  eyepieces  and  bring  the 

specimen in focus.

2. Switch to a low-power objective, select the camera light path and observe 

the monitor image.

3. Loosen the clamping screw @ of the trinocular tube and the clamping 

screw 

2

 of the camera adapter. Then, while observing the monitor image, 

hold the top 

3

 of the camera adapter and turn its bottom | to adjust 

the focus.

} The fixing screw of the camera adapter may not necessarily be located on 

the front side. If you cannot find the fixing screw of the camera adapter on 

the front side, please check either sides or the back side of the camera 

adapter.

4. When the monitor image is focused, tighten the clamping screws @ 

2

 

firmly.

  Adjusting the Centering of the Camera Adapter (Fig. 3)

} The centering adjustment makes it possible to reduce the deviation of 

image between two objectives.

1. Using the Allen screwdriver, loosen the screw marked “LOCK” @ until it 

is projected by 3 or 4 mm from the camera adapter surface.

} Do not loosen the screw too much. Otherwise, it may slip out of the camera 

adapter.

2. Engage the objective with higher power in the light path and move the 

stage to bring the target region of the specimen on the center of the 

monitor image.

3. Engage the objective with lower power in the light path and perform the 

following operation to bring the target region on the center of the monitor 

image.

  (Turn each of the two screws marked “CENTERING” 

2

 using the Allen 

screwdriver so that the target region is brought on the center.)

} To improve the centering accuracy, repeat steps 2 and 3 as required.

4. Tighten the screw marked “LOCK” @ firmly.

  4

  Image Pickup Areas

} The following illustration shows field of view of eyepieces with a FN of 22 and the image pickup areas of the camera 

adapter depending on the image sensor size.

1-inch image sensor

2/3-inch image sensor

1/2-inch image sensor

FN 22

Magnification on 

the monitor

M-02

2

2

Diagonal size of monitor screen

*

Diagonal size of image sensor

*

*

 Note that the actual diagonal lengths of the monitor and image sensor may vary slightly depending on 

manufacturers.
(Reference data for image sensor sizes)
1-inch camera   16.16 mm. 2/3-inch   11 mm. 1/2-inch   8.08 mm. 1/3-inch   6 mm. 1/4-inch   4 mm.

=

Objective power

×

Camera adapter

power (1X)

×

CAUTION

Fig. 2

1

2

4

3

1