8. REPLACEMENT OF LAMP
(1)
Turn off the light source power supply, and unplug the power cord.
(2)
First check to make sure that the lamp is cool enough, then loosen the screw on
the top of the lamp house, and remove the cover.
Size of the fixture screw: M4
Hexagon wrench: 5 mm
(3)
Pull out the lamp, with the internal lever pressed down.
(4)
Pull out the lamp, and remove it from the lamp-holder.
(5)
Install a new lamp, making sure that it is not loose.
※
Though heavily dependent on the usage conditions (the ambient temperature and
humidity, supply voltage, the period of lighting, etc.), the useful life of the lighting
device (lamp house) is estimated to be three (3) years or 15,000 hours, whichever is
shorter
*
, assuming everyday use of about eight (8) hours. The life of the lamp
assumes 50H.
To avoid electric shock and burns, always turn off the power switch,
unplug the cord, and wait until the lamp is cool to the touch for
replacement.
Avoid touching the halogen lamp with your finger or smudging it with
dirt. Doing so can blacken or damage the bulb.
The lamp house needs inspection. Conduct regular inspections
according to
Appendix: Checklist for inspection of the lighting
device (lamp house)
.
Summary of Contents for USPM-RUIII Series
Page 42: ...PV6253 F4E016 USPM RU Ver3 21 004 Software Operation Manual...
Page 89: ...10 Help Menu PV6253 F4E016 47 10 Help Menu 10 1 Version menu Displays the version...
Page 92: ...11 About the Calibration Method for Wavelength Direction PV6253 F4E016 50...
Page 113: ...PV9386 F4E005 1 USPM RU RUIII Installation Instructions Ver 3 21...
Page 130: ...PV9386 F4E005 18 4 In the following screen click the Yes button to complete the uninstallation...