3. Main Window
PV6253-F4E016
13
position where the cursor is located are indicated. When refractive index is displayed on the panel,
wavelength [nm] and refractive index [n] of the cursor position are indicated. When the quantity of
light is displayed on the panel, wavelength [nm] and the value of the quantity of light are indicated.
When reflectance text or the chromaticity diagram is displayed on the panel, the cursor panel
remains inactive.
[<]
and
[>]
move the cursor in steps of one [nm], while
[<<]
and
[>>]
actuate the cursor in steps of
10 [nm]. When the button in the center of the cursor button is pressed, the jump window will
appear. Enter the wavelength of the jumping destination and press the
[OK]
button. A jump will
take place to the specified wavelength. When multiple data are displayed, a jump will take place to
the data selected with the radio button.
*) The cursor can be moved with "SHIFT" + "<-" or "->", or "Ctrl" + "<-" or "->" in the same
way as with
[<<]
,
[<]
,
[>]
, and
[>>]
.
3.5. Buttons
[Background]
:
Executes background measurement.
Similarly, pressing the "F11" key executes background measurement.
[Reference]
: Executes reference measurement.
Similarly, pressing the "F12" key executes reference measurement.
[Measure1] to [Measure10]
: Executes sample measurement.
Similarly, pressing the "F1" to "F10" keys executes sample measurement.
The data line is displayed in the color shown on the button.
[Data name input box]
: Measured data can be assigned any name entering in the input
box located below the Measure button. (The name assigned here is treated as
a test No.)
[Monitor]
: This button is active only when Quantity of Light Graph and Quantity of
Light Text are selected in the [Display] menu. It shows a quantity of light
graph.
[
]
: The radio buttons located on the left of the
[Measure]
buttons are for
selecting data. They indicate whether the data is currently selected or not.
[Data Clear]
: Clears the measured data currently selected with a radio button.
[All Clear]
: Clears all the measured data.
Summary of Contents for USPM-RUIII Series
Page 42: ...PV6253 F4E016 USPM RU Ver3 21 004 Software Operation Manual...
Page 89: ...10 Help Menu PV6253 F4E016 47 10 Help Menu 10 1 Version menu Displays the version...
Page 92: ...11 About the Calibration Method for Wavelength Direction PV6253 F4E016 50...
Page 113: ...PV9386 F4E005 1 USPM RU RUIII Installation Instructions Ver 3 21...
Page 130: ...PV9386 F4E005 18 4 In the following screen click the Yes button to complete the uninstallation...