7. Details of [Standard] Menu
PV6253-F4E016
28
decides OK or NG using the luminous reflectance. Key in the upper-limit value and
lower-limit value. The result will be indicated as Rv: OK/NG in the status bar.
7.1.2. Chromaticity diagram (440nm - 840nm excluded)
When the
[Enable]
check box in [Judge: Chroma] is checked, the program decides OK or
NG using the Chroma C*. Key in the upper-limit value and lower-limit value. The result will
be indicated as C*: OK/NG in the status bar. Moreover, the standard line will be displayed in
a yellow line on the chromaticity diagram.
When the
[Enable]
check box in [Judge: Hue] is checked, the program decides OK or NG
using the hue angle key in the upper-limit value and lower-limit value. The result is indicated
as hkkk: OK/NG in the status bar. Moreover, the standard line will be displayed in a yellow
line on the chromaticity diagram.
[Standard: Chroma]
Check the [Display] check box, and enter the value. Then, the reference value will be
displayed in a blue line.
[Standard: Hue]
Check the [Display] check box, and enter the value. Then, the reference value will be
displayed in a blue line.
7.1.3. Wavelength value
When the
[Enable]
check box in [Wavelength] is checked, the program decides OK or NG
using the wavelength value. Select either to use the peak or the bottom of the reflectance, and
key in the upper-limit value and lower-limit value. The result will be indicated as Wavelength
Value: OK/NG. Moreover, the standard line will be displayed on the chromaticity diagram.
7.2. How to save standards set up
Click
[Save]
to save standard values under an arbitrary file name.
Note: If the standards have been altered, make sure to save them, since the standard file name
only is saved to an configuration file.
7.3. How to open stored standard values
Select
[Load]
, open the standard value data file, and make setting. The name of the file
currently opened is displayed in the status bar of the standard setup window.
Summary of Contents for USPM-RUIII Series
Page 42: ...PV6253 F4E016 USPM RU Ver3 21 004 Software Operation Manual...
Page 89: ...10 Help Menu PV6253 F4E016 47 10 Help Menu 10 1 Version menu Displays the version...
Page 92: ...11 About the Calibration Method for Wavelength Direction PV6253 F4E016 50...
Page 113: ...PV9386 F4E005 1 USPM RU RUIII Installation Instructions Ver 3 21...
Page 130: ...PV9386 F4E005 18 4 In the following screen click the Yes button to complete the uninstallation...