21
Use
5 Use
5.1 Safety information for use
WARNING
Risk of injury to the patient
There is a risk of electric shock if the following conditions apply
simultaneously:
- The telescope is inserted in the patient.
- The light-guide cable is connected to the telescope but not to
the light source.
- The light source connector of the light-guide cable comes in
contact with a conductive surface.
• Completely connect the endoscope before inserting the
endoscope into the patient.
WARNING
Risk of injury to the patient
When energized endoscopes are used in combination with HF
surgical equipment, there are several risks of injury to the patient.
• Follow the instructions for use of all devices connected to the
endoscope for safe application.
CAUTION
Risk of injury to the patient
Light sources emit large amounts of energy. As a result the
connectors of the endoscopic equipment and the distal end of the
endoscope become hot. There is a risk of:
- thermal injury to the patient’s tissue (e.g., from prolonged
exposure to the intense illumination in cavities with small
lumens, or if the distal end of the endoscope is placed into
close proximity to the tissue).
- burns to the patient’s or user’s skin.
- burns or thermal damage to surgical equipment (e.g., surgical
drapes, plastic materials).
• Do not place the endoscopic equipment on the patient’s skin,
on flammable materials, or on heat-sensitive materials.
Summary of Contents for WA4KL100 ULTRA
Page 1: ...INSTRUCTIONS FOR USE TELESCOPE ULTRA WA4KL100 WA4KL130 WA4KL145 WA4KL500 WA4KL530 WA4KL545...
Page 2: ......
Page 23: ...23 Use Only ignite when the laser emission part is visible on the endoscopic image...
Page 52: ......
Page 53: ......
Page 54: ......
Page 55: ......
Page 56: ...W7510750_06 2018 06 07 en Copyright 2018 Olympus Winter Ibe GmbH...