22
Use
• Set the output power of the light source to the minimum level
that is necessary for a sufficient illumination of the target area.
Avoid prolonged exposure to intense illumination.
• Switch off the light source or set the light source to standby
mode whenever the light source is not in use.
CAUTION
Risk of injury to the patient
Light sources emit large amounts of energy. As a result, the
temperatures in front of the light emission surface increase.
There is a risk of thermal injury to the patient’s tissue.
• Do not use high intensity illumination in cavities with small
lumens.
• Do not use high intensity illumination when the distal end of the
endoscope is placed into close proximity to the tissue.
• If the distal end is soiled and impairs proper visualization,
remove the endoscope and clean the distal end.
CAUTION
Risk of injury to the user
The light-guide connector on the endoscope becomes hot. There
is a risk of burns if the endoscope is held by hand for a longer
period.
• Avoid touching the light-guide connector when holding the
endoscope.
CAUTION
Risk of injury to the user
When using a sterile bag during the procedure, the temperature
of the distal end increases significantly. There is a risk of burns to
the patient’s tissue and the sterile bag.
• Switch the light source into standby mode when using the
sterile bag.
NOTICE
Risk of damage to the product
If activated laser equipment touches the energized endoscope,
the energized endoscope may be damaged.
Summary of Contents for WA4KL100 ULTRA
Page 1: ...INSTRUCTIONS FOR USE TELESCOPE ULTRA WA4KL100 WA4KL130 WA4KL145 WA4KL500 WA4KL530 WA4KL545...
Page 2: ......
Page 23: ...23 Use Only ignite when the laser emission part is visible on the endoscopic image...
Page 52: ......
Page 53: ......
Page 54: ......
Page 55: ......
Page 56: ...W7510750_06 2018 06 07 en Copyright 2018 Olympus Winter Ibe GmbH...