24
After use
6 After use
6.1 Removal and disassembly
1. Switch off the light source.
CAUTION
Risk of injury to the user
The connectors on the endoscope become hot. There is a risk of
burns when touching the connectors.
• Let the endoscope cool down after use.
2. Disconnect the light-guide cable from the light source.
3. Disconnect the light-guide cable from the telescope.
4. If applicable, remove the adapters.
5. If applicable, remove the camera head.
If force is required for disassembly, dispose of the product.
6.2 Preparation for reprocessing at the point of use
WARNING
Risk of injury to the patient and/or the medical personnel
If instruments are left in dry condition for a longer period, heavy
debris may dry to the instrument, leading to incrustations that
may be difficult to remove.
• Reprocess this product immediately after use. Do not leave
used instruments overnight before reprocessing.
• Olympus recommends to return the product from the point
of use to the cleaning area in dry condition to avoid protein
fixation caused by detergents or disinfectants.
Immediately after use
• Prepare reusable instruments for later reprocessing directly in
the operating room immediately after use as described in this
section.