33
Reprocessing
7.4.1 Precondition for manual disinfection
The product must be cleaned thoroughly before disinfection.
Thorough cleaning removes both micro-organisms and organic
material. Failure to remove organic material decreases the
effectiveness of the disinfection process.
• To avoid damaging the product, it is recommended to use
commercially available grasping forceps with rubber-padded
jaws to handle the product once it is immersed.
• Grasp the product at its main body.
7.4.2 Manual disinfection procedure
1. Always refer to the latest instructions of the manufacturer of
the solution.
2. Completely immerse the product in Cidex OPA
®
Solution by
Johnson & Johnson. Ensure that all parts are covered.
3. Soak the product in Cidex OPA
®
Solution for as long as
recommended by the manufacturer of the solution, but for at
least 12 minutes.
4. Remove the product from the solution.
5. Rinse the product by completely immersing it in 8 liters of
sterile water for at least 1 minute.
6. Repeat the rinsing step twice. Always use fresh portions of
sterile water for each rinse.
7. Dry the product.
7.5 Automated cleaning and thermal disinfection
WARNING
Risk of injury to the patient and/or the medical personnel
There is an infection control risk when using insufficient
detergents.
• Only use detergents which have been validated for the product.
Summary of Contents for WA4KL100 ULTRA
Page 1: ...INSTRUCTIONS FOR USE TELESCOPE ULTRA WA4KL100 WA4KL130 WA4KL145 WA4KL500 WA4KL530 WA4KL545...
Page 2: ......