35
Reprocessing
Automated cleaning of the product has been validated with the
following detergents:
- Neodisher MediClean forte manufactured by Chemische Fabrik
Dr. Weigert GmbH und Co. KG
- 0.5% solution
-
pH value: approx. 10.3
- Neodisher Z manufactured by Chemische Fabrik Dr. Weigert
GmbH und Co. KG
- 0.1% solution
-
pH value: approx. 1.2
NOTICE
Risk of damage to the product
Detergents or tap water may contain various aggressive
compounds that can corrode the product.
• To remove residues, rinse the product thoroughly with
deionized water.
• Do not use tap water for rinsing because it might be
chlorinated.
7.5.1 Preparation for automatic cleaning and thermal disinfection
•
Use adequate instrument trays to fix the product.
•
Make sure that the product has been securely fixed to the
trays or baskets of the washer-disinfector. Make sure that
instruments do not touch each other.
• Refer to the instructions for use of the instrument tray and the
washer-disinfector.
• Make sure that instruments are placed in the washer-disinfector
so that they can be appropriately cleaned and disinfected.
• Do not overload the washer-disinfector.
7.5.2 Selection of programs
Select the program according to the following characteristics:
- The program must at least reach an A0 value of 3000.
-
The temperature of the first inlet water must not exceed 20 °C.
Summary of Contents for WA4KL100 ULTRA
Page 1: ...INSTRUCTIONS FOR USE TELESCOPE ULTRA WA4KL100 WA4KL130 WA4KL145 WA4KL500 WA4KL530 WA4KL545...
Page 2: ......
Page 23: ...23 Use Only ignite when the laser emission part is visible on the endoscopic image...
Page 52: ......
Page 53: ......
Page 54: ......
Page 55: ......
Page 56: ...W7510750_06 2018 06 07 en Copyright 2018 Olympus Winter Ibe GmbH...