39
Reprocessing
3. Press the telescope into the notches.
4. Close the instrument tray.
5. Seal the instrument tray in packaging appropriate for
sterilization according to local guidelines and your hospital’s
protocol.
Use of instrument trays
• Do not stack the instrument trays or load them on their side
edges during steam sterilization.
Steam sterilization conditions
Steam sterilization of the product has been validated with
fractionated prevacuum at 132 °C for 3 min.
Other possible steam sterilization parameters are:
Temperature ................................................... >132 °C (269.6 °F)
Holding time ........................................................................ >3 min
NOTICE
Risk of damage to the product
Too high temperatures may damage the product.
• Do not exceed a temperature of 138 °C. Otherwise, damage
may occur.
Summary of Contents for WA4KL100 ULTRA
Page 1: ...INSTRUCTIONS FOR USE TELESCOPE ULTRA WA4KL100 WA4KL130 WA4KL145 WA4KL500 WA4KL530 WA4KL545...
Page 2: ......
Page 23: ...23 Use Only ignite when the laser emission part is visible on the endoscopic image...
Page 52: ......
Page 53: ......
Page 54: ......
Page 55: ......
Page 56: ...W7510750_06 2018 06 07 en Copyright 2018 Olympus Winter Ibe GmbH...