Reprocessing
• After steam sterilization, let the instrument cool down gradually
to room temperature without additional cooling. Sudden
changes in temperature may damage the product.
Steam sterilization procedure
For steam sterilization, Olympus recommends a drying time of at
least 15–30 minutes.
• Choose an appropriate drying time, depending on the
autoclave.
• Refer to the autoclave’s instructions for use.
• Be careful when unloading the autoclave, the contents may be
hot.
• Make sure that the sterile packaging of the product is not
damaged.
•
Resterilize the product if 1 of the following applies:
- the sealing has been opened
- the packaging is wet
- the packaging is damaged