42
Reprocessing
• When sterilizing the endoscope in the plastic instrument tray,
do not use STERRAD
®
.
Preparing the product for STERRAD
®
plasma sterilization
• Place the product in sterilizing pouches or containers (any
container/pouches must be certified by its manufacturer for
STERRAD
®
use).
• Only use sterilization accessories that have been cleared
by the local authorities for the use within the respective
STERRAD
®
system.
• Refer to the instructions for use of the sterilizer.
STERRAD
®
systems
Different STERRAD
®
systems with different cycles exist in the
market. The information provided on STERRAD
®
sterilization in
this document applies to the following STERRAD
®
systems and
cycles:
- STERRAD
®
50: only 1 cycle
- STERRAD
®
100S: short cycle and long cycle
- STERRAD
®
200: short cycle and long cycle
- STERRAD
®
NX™: standard cycle and advanced cycle
- STERRAD
®
100NX™: standard cycle and express cycle
7.9
Verified sterilization methods
The sterilization methods described in this section have been
verified for material compatibility only.
Verified for material compatibility means that according to
current knowledge the reprocessing method and/or agent does
not negatively affect materials or functional performance of the
product.
Verified for material compatibility does not mean that
microbiological efficacy can be guaranteed.
Summary of Contents for WA4KL100 ULTRA
Page 1: ...INSTRUCTIONS FOR USE TELESCOPE ULTRA WA4KL100 WA4KL130 WA4KL145 WA4KL500 WA4KL530 WA4KL545...
Page 2: ......
Page 23: ...23 Use Only ignite when the laser emission part is visible on the endoscopic image...
Page 52: ......
Page 53: ......
Page 54: ......
Page 55: ......
Page 56: ...W7510750_06 2018 06 07 en Copyright 2018 Olympus Winter Ibe GmbH...