Compatible equipment
9 Compatible equipment
Olympus recommends the use of equipment as listed in this
chapter. If combinations are used that are not listed in this
chapter, the user takes the full responsibility.
Future equipment may also be compatible. For more information,
contact an Olympus representative.
Some of the products listed in this chapter may not be available
in all sales territories.
CAUTION
Risk of injury to the patient and/or the user
When using endoscopes and light-guide cables with an
inappropriate combination of the maximum insertion portion width
and fiber bundle diameter, the connectors may become hot.
• Only use endoscopes and light-guide cables with the
appropriate combination of the maximum insertion portion
width and fiber bundle diameter.
• For the appropriate maximum insertion portion width of the
telescope, refer to the chapter “Technical data” on page 50.
•
For the appropriate fiber bundle diameter, refer to the
respective instructions for use of the light-guide cable.
Light-guide cables
Olympus recommends the use of an Olympus light-guide
cable and an Olympus light source. Only this combination will
guarantee optimum illumination of the endoscopic image and
excellent color reproduction.
Using endoscopes in combination with light-guide cables of other
manufacturers may lead to an increase in temperature of the
connectors.
Summary of Contents for WA4KL100 ULTRA
Page 1: ...INSTRUCTIONS FOR USE TELESCOPE ULTRA WA4KL100 WA4KL130 WA4KL145 WA4KL500 WA4KL530 WA4KL545...
Page 2: ......
Page 23: ...23 Use Only ignite when the laser emission part is visible on the endoscopic image...
Page 52: ......
Page 53: ......
Page 54: ......
Page 55: ......
Page 56: ...W7510750_06 2018 06 07 en Copyright 2018 Olympus Winter Ibe GmbH...