Page 11 of 76
Issue 20 : July 2012
WM--P1 SERIES WORKSTATIONS
Potential
equalisation
Terminals fitted
2 at input, 2 x 6 at output
Standards compliance
The Separation Transformer conforms to the relevant
requirements of
EN IEC 60601--1/EN IEC 60601--1--1:2001, Annex EEE/UL
60601--1 and CAN/CSA Std. No. C22.2 No. 601.1--M90
Classification
Type of protection against electric
shock.
Class I equipment.
Classification
Degree of protection against
explosion
NONE
-- The Separation transformer should not be used
within the zone of risk of flammable gases
Degree of protection against ingress
of water.
IPXO as defined in EN IEC 60529:1992, ordinary equipment.
Mode of operation
Continuous
Ambient temperature
Storage: --40
C to 70
C (--40
F to 158
F)
Environmental
Ambient temperature
Operation: 10
C to 40
C (50
F to 104
F)
Environmental
conditions
Atmospheric pressure
Storage/Operation: 70--106 kPa (10 3--15 4lbf/in2)
conditions
Atmospheric pressure
Storage/Operation: 70--106 kPa (10.3--15.4lbf/in2)
Relative humidity
95% maximum @ 40
C (104
F) non--condensing
Regulatory status
220--240V units are CE marked in
accordance with Directive 93/42/EEC of
14 June 1993 concerning medical
devices, as an accessory to a medical
device, Class I. Year of manufacture is
given in first two digits of serial number.
End of Life
European Economic Area (EEA)
In accordance with European Directive
2002/96/EC on Waste Electrical and
Electronic Equipment, this symbol
indicates that the product must not be
disposed of as unsorted municipal waste,
but should be collected separately. Refer
to your local Olympus distributor for return
and/or collection systems available in your
country.
The Olympus WM--P1 series Mobile Workstation and Separation Transformer is manufactured in the
UK by KeyMed (Medical & Industrial Equipment) Ltd.