Page 25 of 76
Issue 20 : July 2012
WM--P1 SERIES WORKSTATIONS
6.12 Scope pole clamp P1 series
(1)
Refer to Figure 6-1. Blanking caps : Cable port (item 17), Monitor (item 18) and Scope pole (item 19) are a push fit and
so require to be carefully prised out of position.
(2)
Refer to Figure 6-9 Unscrew Handknob (item 6) to release Cable tidy (item 29).
(3)
Remove the top tray, refer to section 6.8 for WM--NP1,section 6.9 for WM--WP1 and section 6.10 for WM--DP1.
(4)
Refer to Figure 6-9 Use a pozi screwdriver to release screw (item 41) with washer (item 48) and remove Scope Pole
Clamp Assy (item 46).
(5)
Refit in reverse order ensuring that Screw (item 41) is tightend to 2.5Nm using a No2 pozi drive tool bit in powered torque
tool.
(6)
Refit remaining parts in reverse order.
FIGURE 6-9
SCOPE POLE CLAMP, SCOPE POLE
RUNNERS & NUT STRIPS
48
8
7
6
28
2
12
4
13
8
41
46
2
29
6.13 Runner- scope pole P1 series
(1)
Refer to section 5.11 to remove scope pole clamp P1 series.
(2)
Lift each runner from it’s slot in the side column.
(3)
Re fit in reverse order.
6.14 Nutstrip P1 series
(1)
Remove the top tray refer to section 5.8 for WM--NP1, section 5.9 for WM--WP1 and section 5.10 for WM--DP1
(2)
Refer to Figure 6-9 and remove Nutstips (item 12) noting which side of the nutstrip is facing outward.
(3)
Replace/refit in reverse order.