Troubleshooting Guide 

 

Optical Problems

 

Problem

 

Cause

 

Solution

 

Darkness at the 
periphery or uneven 
brightness in the 
field of view 

Revolving nosepiece not in click 
stop position 

Revolve the nosepiece to click-stop 
position by swinging the objective 
correctly into the optical path 

The light source of the bulb is not at 
the center 

Adjust the position of the bulb 

Dirt or dust on the 
view 

Dirt or dust on the lens eyepiece, 
condenser, objective, collector lens 
or specimen 

Clean the lens with a camera 
cleaning kit 

Poor image quality 

No slide cover attached to the slide 

Attach a 0.17mm slide cover 

Slide cover is too thick or thin 

Use a slide cover of the appropriate 
thickness (0.17mm) 

Slide may be upside down 

Turn slide over so the cover-glass 
faces up 

Immersion oil is on a dry objective 
(especially the 40x) 

Check the objectives, clean if 
necessary 

No immersion oil used with 100x 
objective 

Use immersion oil 

Air bubbles in immersion oil 

Remove bubbles 

Condenser aperture is closed or 
open too much 

Open or close properly 

Condenser is positioned too low 

Position the condenser upward 

 

 

Electrical Problems

 

Problem

 

Cause

 

Solution

 

Lamp does not light 
when switched on 

No electrical power   

Check power cord connection   

Lamp bulb burnt out   

Replace bulb   

Fuse blown out 

Replace fuse 

 

 

 

 

 

Summary of Contents for M8244

Page 1: ...User Manual Dual Head Binocular Compound Microscope Model M8244 MicroscopeNet com...

Page 2: ...Table of Contents i Caution 1 ii Care and Maintenance 2 1 Component Illustration 3 2 Installation 4 3 Operation 6 4 Specifications 8 5 Troubleshooting Guide 9...

Page 3: ...ortant the lamp lamp housing and adjacent parts will become very hot Do not touch these parts until they have completely cooled Never attempt to handle a hot halogen bulb 6 All electrical connectors p...

Page 4: ...he optical surface use a soft cloth or cotton swab dampened with a lens cleaning solution available at camera stores All optical lenses should be swabbed using a circular motion A small amount of abso...

Page 5: ...8 Mechanical Stage 9 Condenser 10 Condenser Focus Knob 11 Light Collector 12 Base 13 Viewing Head V 14 Thumb Screw V 15 Thumb Screw H 16 Viewing Head Block 17 Thumb Screw 18 Body 19 Tension Collar 20...

Page 6: ...eeding higher magnification objective Note When changing the objective magnification rotate the objective nosepiece until you hear a click sound This ensures the objective is centered in the optical l...

Page 7: ...pieces off and be certain that the head is securely locked by the thumb screw 2 7 Installing the mirror optional your model may not have one 1 Unplug the power cord 2 Screw off the light collector Len...

Page 8: ...highest position without contacting the 100x objective then tighten the upper limit mechanical stage lever 3 4 Condenser focusing Turn the condenser focus knob 10 to raise or lower the condenser The...

Page 9: ...harp image is obtained 3 Since both sides are adjustable you may also do the above in the opposite way in other words left eye first and right eye second 3 8 Adjusting tension knob The tightness of th...

Page 10: ...nt Focus Mechanism Coaxial coarse and fine focusing knobs on both sides w limit lock Minimum fine focusing adjustment at 0 002mm range 28 mm Stage Large double layer mechanical stage Dimension 5 1 2 x...

Page 11: ...quality No slide cover attached to the slide Attach a 0 17mm slide cover Slide cover is too thick or thin Use a slide cover of the appropriate thickness 0 17mm Slide may be upside down Turn slide over...

Page 12: ...nsity control dial Aperture diaphragm closed too far Open to the proper setting Condenser position too low Position the condenser upward Mechanical Problems Problem Cause Solution Image will not focus...

Reviews: