Troubleshooting Guide 

GENERAL PROBLEMS

 

Problem

 

Cause

 

Solution

 

Lamp does not light 
when switched on 

No electrical power   

Check power cord connection   

LED or power unit dead   

Replace LED light 

Fuse blown out 

Replace fuse 

Darkness at the 
periphery or uneven 
brightness in the field 
of view 

Revolving nosepiece not in click stop 
position 

Revolve the nosepiece to click-stop 
position by swinging the objective 
correctly into the optical path 

The light source of the lamp is not at 
the center 

Adjust the position of the lamp 

Dirt or dust on the 
view 

Dirt or dust on the lens eyepiece, 
condenser, objective, collector lens or 
specimen 

Clean the lens with a camera cleaning kit 

Poor image quality or 
not able to get 
focused image 

No slide cover attached to the slide 

Attach a 0.17mm slide cover 

Slide cover is too thick or thin 

Use a slide cover of the appropriate 
thickness (0.17mm) 

Slide may be upside down (specimen 
at the bottom) 

Turn slide over so the cover-glass faces 
up 

Diopter adjustment is not set properly 

Readjust the diopter settings 

Immersion oil is on a dry objective 
(especially the 40X) 

Check the objectives, clean if necessary 

No immersion oil used with 100X 
objective 

Use immersion oil 

Air bubbles in immersion oil 

Remove bubbles 

Condenser aperture is closed or open 
too much 

Open or close properly 

Condenser is positioned too low 

Position the condenser upward 

Specimen rises from stage surface 

Secure the specimen in the slide holder 

Blue filter not used 

Use daylight blue filter 

Lamp intensity is too high or low 

Adjust the light intensity by rotating the 
intensity control dial 

Slippage of focus 
when using the 
coarse focusing knob 
Fine focus is 
ineffective 

Tension adjustment is set too low 

Increase the tension on the focusing 
knobs 

Tension adjustment is set too high 

Loosen the tension on the focusing 
knobs 

 

Summary of Contents for M827TL Series

Page 1: ...User Manual Binocular Compound LED Microscope Model M827TL Series MicroscopeNet com...

Page 2: ...tion 1 ii Care and Maintenance 2 2 Components Illustration 3 3 Installation 4 4 Operation 6 5 Specifications 10 6 Optional Parts 11 7 Troubleshooting Guide 13 8 Darkfield Condenser Installation and Op...

Page 3: ...of direct sunlight high temperature or humidity and dusty environments Ensure that the microscope is located on a smooth level and firm surface 4 If any specimen solutions or other liquids splash onto...

Page 4: ...ng an excessive amount of solvents as this may cause problems with optical coatings or cemented optics or the flowing solvent may pick up grease making cleaning more difficult Oil immersion objectives...

Page 5: ...Screw 19 Focus Stop Lever 4 Nosepiece 12 Microscope Body 20 Condenser Lock Thumb Knob 5 Objective 13 Focus Tension Ring 21 Condenser Focus Knob 6 Slide Holder 14 Coarse Focus Knob 22 Abbe Condenser 7...

Page 6: ...mechanical stage is at its lowest position 2 Turn the caps counter clockwise to remove them from the nosepiece 3 Take the objectives out from the plastic cases and turn each one clock wise into the h...

Page 7: ...take the eyepieces off and be certain that the head is securely locked by the thumb screw 2 6 Installing the mirror optional may not included in your package 1 Turn off the power switch and disconnec...

Page 8: ...d into the power socket on the microscope and connect it to the power outlet 2 Turn on the power switch 3 Rotate the brightness intensity dial to increase or decrease the brightness Caution A diffusio...

Page 9: ...get a good focused image you may need to combine the focus knob adjustment and interpupillary distance adjustment along with eyepiece diopter adjustment stated in 3 3 and 3 4 5 You may now switch to...

Page 10: ...s first 3 7 Adjusting condenser 1 Turn the condenser focus knob to raise or lower the condenser 2 Raise the condenser when using high power objectives and lower it when using low power objectives Note...

Page 11: ...by following the procedures in 3 5 2 Install the camera by following the procedures in 2 7 3 Open image observing sofeware to examine more details see camera s manual 4 You also can capture images or...

Page 12: ...and filter holder Rack and pinion adjustment Focus Mechanism Coaxial coarse and fine focusing knobs on both sides with focus stop Minimum fine focusing adjustment at 0 002mm range 28mm Mechanical Stag...

Page 13: ...d separately 1 Cameras Model Resolution Operating System Software A1502 640 x 480 0 3MP MS Windows Mac OS X 10 6 10 8 Included A3513U 1280 x 1024 1 3MP MS Windows Mac OS 10 8 and up Linux 2 6 or above...

Page 14: ...ze diameter A191 Dry 0 7 0 9 37mm A191BD Oil 1 36 1 25 37mm A191BOIL Oil 1 36 1 25 Plan 100X 1 25 oil 0 5 160 0 17 spring w iris diaphragm 37mm 3 Carrying Case Model Dimension Net Weight A187V 9 7 8 x...

Page 15: ...is too thick or thin Use a slide cover of the appropriate thickness 0 17mm Slide may be upside down specimen at the bottom Turn slide over so the cover glass faces up Diopter adjustment is not set pr...

Page 16: ...e condenser is not in the right position Lower or raise the condenser slightly to the position that the specimen is the brightest A191BOIL A191BD only Make sure the oil contact the bottom of the slide...

Page 17: ...ondenser up or down slightly to get the best darkfield viewing Note The dry darkfield condenser is used with the dry objectives only The dry darkfield condenser works with the 4X 10X 40X objectives Th...

Page 18: ...jective you will need to drop the immersion oil and adjust the iris ring Fig 15 to get proper brightness and contrast of view field 8 After observing the specimen with 100X objective use the lens clea...

Reviews: