Troubleshooting Guide 

Problem

 

Cause

 

Solution

 

Lamp does not light 
when switched on 

No electrical power   

Check power cord connection   

LED or power unit dead 

Contact seller for service   

Darkness at the 
periphery 

Revolving nosepiece not in click stop position 

Revolve the nosepiece to click-stop 
position by swinging the objective 
correctly into the optical path 

Uneven brightness 
in the field of view 

The diffusion filter is not in the light path 

Check the diffusion filter and make sure 
it’s in the light path 

Insufficient 
brightness 

Diffusion filter used when using 100X objective 

Swing out the diffusion filter when using 
100X objective 

Dirt or dust on the 
view 

Dirt or dust on the lens eyepiece, condenser, 
objective, collector lens or specimen 

Clean the lens with a lens cleaning 
paper 

Dirt or dust on the specimen 

Clean the specimen 

Poor image quality 
or not able to get 
focused image 

No slide cover attached to the slide 

Attach a 0.17mm slide cover 

Slide cover is too thick or thin 

Use a slide cover of the appropriate 
thickness (0.17mm) 

Slide may be upside down (specimen at the 
bottom) 

Turn slide over so the cover-glass 
faces up 

Diopter adjustment is not set properly 

Readjust the diopter settings 

Immersion oil is on a dry objective (especially 
the 40X) 

Check the objectives, clean if 
necessary 

No immersion oil used with 100X objective 

Use immersion oil 

Air bubbles in immersion oil 

Remove bubbles 

Specimen rises from stage surface 

Secure the specimen in the slide holder 

Condenser aperture is closed or open too much 

Open or close properly 

Condenser is positioned too low or too high 

Move condenser upper or lower 

Lamp intensity is too high or low 

Adjust the light intensity by rotating the 
intensity control dial 

Image is black 
screen through 
camera 

No enough lighting or driver not installed 
properly 

Check the camera manual 

Image is totally 
white through 
camera   

The light is too bright 

Manually adjust the exposure in the 
software. See manual for details. 

Summary of Contents for MD82ES10

Page 1: ...User Manual Digital Compound Binocular LED Microscope Model MD82ES10 MicroscopeNet com...

Page 2: ...tents i Caution 1 ii Care and Maintenance 2 1 Components Illustration 3 2 Installation 4 3 Operation 4 4 Specifications 8 5 Troubleshooting Guide 9 6 Darkfield Condenser Installation and Operation Ins...

Page 3: ...ments Ensure the microscope is located on a smooth level and firm surface 4 If any specimen solutions or other liquids splash onto the stage objective or any other component disconnect the power cord...

Page 4: ...ng an excessive amount of solvents as this may cause problems with optical coatings or cemented optics or the flowing solvent may pick up grease making cleaning more difficult Oil immersion objectives...

Page 5: ...ve 17 Power Switch 6 Slide Holder 18 Brightness Intensity Dial 7 Mechanical Stage 19 Abbe Condenser 8 Light Collector 20 Condenser Lock Thumb Screw 9 Microscope Base 21 Condenser Control Knob 10 Viewi...

Page 6: ...n 6 Dry Darkfield Condenser Installation Operation Instruction 3 Operation 3 1 Adjusting illumination 1 Plug the power cord into the power socket on the microscope and connect it to the power outlet 2...

Page 7: ...same specimen into clear sharp focus by adjusting the diopter ring on left eyepiece tube only Do not use focus knobs at this step 5 Since both sides are adjustable you may also do the above in the op...

Page 8: ...When you use the 100X objective to observe the specimen you have to apply the immersion oil on the top of the slide cover When you apply the immersion oil with the 100X objective do not let the immers...

Page 9: ...back of viewing head and the other end to the computer 3 Turn on the computer install the camera following the manual in the mini CD 4 Open image observing software to examine more details see camera...

Page 10: ...Condenser Abbe NA 1 25 w iris diaphragm and diffusion filter Rack and pinion adjustment Focus Mechanism Coaxial coarse and fine focusing knobs on both sides with rack stop Mechanical Stage Double laye...

Page 11: ...get focused image No slide cover attached to the slide Attach a 0 17mm slide cover Slide cover is too thick or thin Use a slide cover of the appropriate thickness 0 17mm Slide may be upside down speci...

Page 12: ...se the bolt to lock it when needed Note When raising the mechanical stage do not make contact with the objective 2 Centering the dry darkfield condenser Turn the 4X objective to the light path Turn th...

Reviews: