Darkfield Condenser Installation and 
Operation Instructions 

1)  Mounting the dry darkfield condenser 

  Rotate the nosepiece and set the 4x objective in 

position. 

  Loosen and raise the focusing rack stop screw to 

the highest position. 

  Turn the coarse focus knob to raise the 

mechanical stage to the highest position without 
contacting the 4X objective. 

  Turn the condenser control knob to raise the 

condenser to the highest position. 

  Loosen the condenser lock thumb screw, pull down the condenser and take it off, 

insert the dry darkfield condenser in, and tighten the condenser lock thumb 
screw (See 

Fig. 7

). 

  Re-adjust the condenser height to its normal position. 

  Re-adjust the mechanical stage height to its normal position. 

  Lower the focusing rack stop screw back and use the bolt to lock it when 

needed. 

Note:   

When raising the mechanical stage, do not make contact with the objective. 

2)  Centering the dry darkfield condenser 

  Turn the 4X objective to the light path. 

  Turn the condenser focus knob to lower the 

condenser till a dark spot showed in the 
viewing field as shown in 

Fig. 8

 

(a)

.

 

  Turn the condenser translational centering 

screws to move the dark spot to the center 
as shown 

in 

Fig. 8

 

(b)

3)  Place the slide on the stage. 
4)  Raise the condenser all the way to the top and then lower it a little bit. 
5)  Follow the procedures in this manual to bring the sample on slide in focus and 

observe. 

6)  Move the condenser up or down slightly to get the best darkfield viewing. 

Note

 

  The dry darkfield condenser is used with the dry objectives only. 

  The dry darkfield condenser works with the 4X, 10X, 40X objectives. 

 

The dry darkfield condenser does not work with the 100X oil immersion 
objective.

 

 

 

     

(a)

                             

  (b)

 

Fig. 7 

Fig. 8 

Condenser Lock   

Thumb Screw 

Centering Screws 

Condenser Holder 

Summary of Contents for MD82ES10

Page 1: ...User Manual Digital Compound Binocular LED Microscope Model MD82ES10 MicroscopeNet com...

Page 2: ...tents i Caution 1 ii Care and Maintenance 2 1 Components Illustration 3 2 Installation 4 3 Operation 4 4 Specifications 8 5 Troubleshooting Guide 9 6 Darkfield Condenser Installation and Operation Ins...

Page 3: ...ments Ensure the microscope is located on a smooth level and firm surface 4 If any specimen solutions or other liquids splash onto the stage objective or any other component disconnect the power cord...

Page 4: ...ng an excessive amount of solvents as this may cause problems with optical coatings or cemented optics or the flowing solvent may pick up grease making cleaning more difficult Oil immersion objectives...

Page 5: ...ve 17 Power Switch 6 Slide Holder 18 Brightness Intensity Dial 7 Mechanical Stage 19 Abbe Condenser 8 Light Collector 20 Condenser Lock Thumb Screw 9 Microscope Base 21 Condenser Control Knob 10 Viewi...

Page 6: ...n 6 Dry Darkfield Condenser Installation Operation Instruction 3 Operation 3 1 Adjusting illumination 1 Plug the power cord into the power socket on the microscope and connect it to the power outlet 2...

Page 7: ...same specimen into clear sharp focus by adjusting the diopter ring on left eyepiece tube only Do not use focus knobs at this step 5 Since both sides are adjustable you may also do the above in the op...

Page 8: ...When you use the 100X objective to observe the specimen you have to apply the immersion oil on the top of the slide cover When you apply the immersion oil with the 100X objective do not let the immers...

Page 9: ...back of viewing head and the other end to the computer 3 Turn on the computer install the camera following the manual in the mini CD 4 Open image observing software to examine more details see camera...

Page 10: ...Condenser Abbe NA 1 25 w iris diaphragm and diffusion filter Rack and pinion adjustment Focus Mechanism Coaxial coarse and fine focusing knobs on both sides with rack stop Mechanical Stage Double laye...

Page 11: ...get focused image No slide cover attached to the slide Attach a 0 17mm slide cover Slide cover is too thick or thin Use a slide cover of the appropriate thickness 0 17mm Slide may be upside down speci...

Page 12: ...se the bolt to lock it when needed Note When raising the mechanical stage do not make contact with the objective 2 Centering the dry darkfield condenser Turn the 4X objective to the light path Turn th...

Reviews: