3.2  Placing specimen 

1)  Place the slide on the mechanical stage.   
2)  Use the slide holder to gently secure the slide.   
3)  Turn the X and Y stage moving knobs to position the specimen in the center of 

viewing field. 

Caution:

   

Be sure not to allow an objective to touch a specimen slide when changing objectives.

 

3.3  Adjusting interpupillary distance 

While observing with both eyes, hold the left and 
right eyepiece tubes and slide the tubes in and 
out. The interpupillary distance is correct when 
the left and right fields of view converge 
completely into one image. 

3.4  Adjusting eyepiece diopter 

1)  Rotate the 10X objective into position. 
2)  Rotate the diopter ring on the right eyepiece 

tube until its numerical value is the same as 
your interpupillary distance, for example, 

70

 in 

the right figure (See 

Fig. 2

). 

3)  Close your left eye and bring the specimen into 

focus following the focusing procedures in 3.5. 

4)  Close your right eye and bring the same 

specimen into clear sharp focus by adjusting 
the diopter ring on left eyepiece tube only. Do 
not use focus knobs at this step. 

5)  Since both sides are adjustable, you may also do the above in the opposite way, in 

other words, left eye first and right eye second. 

3.5  Focusing 

1)  With the 10X objective in position, raise the mechanical stage using the coarse 

focus knob until the specimen is close to the objective 

2)  Turn the coarse focus knob until the specimen is in focus. Use the fine focus knob 

to obtain a sharp image. You may now switch to another magnification objective. 

3)  To get a good focused image, you may need to combine the focus knob adjustment 

and interpupillary distance adjustment, along with eyepiece 
diopter adjustment stated in 

3.3

 and 

3.4

Tips: 

 

T

o prevent your specimen slide from making contact with an 

objective, turn the 100X objective in position and adjust the 
focusing rack stop screw (

Fig. 3

) so that the 100X objective 

will not contact the specimen while the stage is adjusted to 
its highest position. Give the stage a tiny extra moving 
space to ensure the objective can be focused every time. 
 
 

70 

Fig. 2 

Fig. 3 

Focusing Rack Stop 

Screw 

Summary of Contents for MD82ES10

Page 1: ...User Manual Digital Compound Binocular LED Microscope Model MD82ES10 MicroscopeNet com...

Page 2: ...tents i Caution 1 ii Care and Maintenance 2 1 Components Illustration 3 2 Installation 4 3 Operation 4 4 Specifications 8 5 Troubleshooting Guide 9 6 Darkfield Condenser Installation and Operation Ins...

Page 3: ...ments Ensure the microscope is located on a smooth level and firm surface 4 If any specimen solutions or other liquids splash onto the stage objective or any other component disconnect the power cord...

Page 4: ...ng an excessive amount of solvents as this may cause problems with optical coatings or cemented optics or the flowing solvent may pick up grease making cleaning more difficult Oil immersion objectives...

Page 5: ...ve 17 Power Switch 6 Slide Holder 18 Brightness Intensity Dial 7 Mechanical Stage 19 Abbe Condenser 8 Light Collector 20 Condenser Lock Thumb Screw 9 Microscope Base 21 Condenser Control Knob 10 Viewi...

Page 6: ...n 6 Dry Darkfield Condenser Installation Operation Instruction 3 Operation 3 1 Adjusting illumination 1 Plug the power cord into the power socket on the microscope and connect it to the power outlet 2...

Page 7: ...same specimen into clear sharp focus by adjusting the diopter ring on left eyepiece tube only Do not use focus knobs at this step 5 Since both sides are adjustable you may also do the above in the op...

Page 8: ...When you use the 100X objective to observe the specimen you have to apply the immersion oil on the top of the slide cover When you apply the immersion oil with the 100X objective do not let the immers...

Page 9: ...back of viewing head and the other end to the computer 3 Turn on the computer install the camera following the manual in the mini CD 4 Open image observing software to examine more details see camera...

Page 10: ...Condenser Abbe NA 1 25 w iris diaphragm and diffusion filter Rack and pinion adjustment Focus Mechanism Coaxial coarse and fine focusing knobs on both sides with rack stop Mechanical Stage Double laye...

Page 11: ...get focused image No slide cover attached to the slide Attach a 0 17mm slide cover Slide cover is too thick or thin Use a slide cover of the appropriate thickness 0 17mm Slide may be upside down speci...

Page 12: ...se the bolt to lock it when needed Note When raising the mechanical stage do not make contact with the objective 2 Centering the dry darkfield condenser Turn the 4X objective to the light path Turn th...

Reviews: