background image

Entering Code 66 will overwrite all user settings with

the factory settings. The meter will display 

rESEt

and then

return to 

CodE 00

. Press the 

SEL

button to exit the module. 

10

6.2  MODULE 2 - S

ECONDARY

F

UNCTION

P

ARAMETERS

(

2-SEC

)

PARAMETER MENU

MIN DISPLAY ENABLE

USER INPUT ASSIGNMENT

Select the value(s) to which the User Input Function is assigned. The User

Input Assignment only applies if a selection of reset, display hold, or print and
reset is selected in the User Input Function menu.

MODE

DISPLAY

No Function

NO

DESCRIPTION

User Input disabled.

Program Mode Lock-out

P-Loc

Zero Input 
(Edge triggered)

ZErO

Zero the Input Display value causing
Display Reading to be Offset.

COLOr

d-SEL

d-LEV

d-HLd

rESEt

Resets the assigned value(s) to the
current input value.

HI-LO 

HI

dSP

LO

See Programming Mode Access chart
(Module 3).

YES

NO

MODE

DESCRIPTION

DISPLAY

Setpoint 1 and 2 Reset

Setpoint 1 Reset

Setpoint 2 Reset

Print and Reset

Reset both setpoint 1 and 2 outputs.

Resets setpoint 1 output.

Resets setpoint 2 output.

Same as Print Request followed by a
momentary reset of the assigned value(s).

rSt12

rSt-1

rSt-2

P-r5t

Print Request

Serial transmit of the active parameters
selected in the Print Options menu
(Module 5).

Print

ª

«

2.0

HI-t

ª

«

NO

LO-En

ª

«

NO

FCS

MAX CAPTURE DELAY TIME

When the Input Display is above the present MAX value for the entered

delay time, the meter will capture that display value as the new MAX reading.
A delay time helps to avoid false captures of sudden short spikes.

ª

«

2.0

LO-t

MIN CAPTURE DELAY TIME

When the Input Display is below the present MIN value for the entered delay

time, the meter will capture that display value as the new MIN reading. A delay
time helps to avoid false captures of sudden short spikes.

0.0

to

999.9

sec.

USER INPUT FUNCTION

ª

«

NO

USrIN

ª

«

dSP

U-ASN

ª

«

NO

HI-En

MAX DISPLAY ENABLE

0.0

to

999.9

sec.

Select 

yES

to perform either of the Factory Service Operations shown below.

FACTORY SERVICE OPERATIONS

yES

NO

YES

NO

Enables the Maximum Display Capture capability.

Enables the Minimum Display Capture capability.

The unit uses stored current calibration values to

provide accurate current measurements. Over time, the
electrical characteristics of the components inside the unit
will slowly change with the result that the stored

calibration values no longer accurately define the input circuit. For most
applications, recalibration every 1 to 2 years should be sufficient.

Calibration of the unit involves a current calibration which should only be

performed by individuals experienced in calibrating electronic equipment.
Allow 30 minute warm up before performing any calibration related procedure.
The following procedures should be performed at an ambient temperature of 15
to 35 °C (59 to 95 °F).

CAUTION: The accuracy of the calibration equipment will directly affect the

accuracy of the unit.

Current Calibration

1. Connect the negative lead of a precision DC current source with an accuracy

of 0.01% or better to the COMM terminal. Leave the positive lead of the DC
current source unconnected.

2. With the display at 

CodE 48

, press and hold the 

SEL

button for 2 seconds. Unit

will display 

CAL NO

3. Press the 

RST

button to select the range to be calibrated.

4. Press the 

SEL

button. Display reads 0.0A

5. With the positive lead of the DC current source unconnected, press 

SEL

.

Display reads 

CALC

for about 8 seconds.

6. When the display reads the selected range, connect the positive lead of the DC

current source to INP+ and apply full-scale input signal for the range. (Note:
For 200 mA range, apply 100 mA as indicated on the display.)

7. Repeat steps 3 through 6 for each input range to be calibrated. When display

reads 

CAL NO

, press the 

SEL

button to exit calibration.

CALIBRATION

ª

«

48

CodE

ª

«

66

CodE

RESTORE FACTORY DEFAULT SETTINGS

Backlight Color 
(Edge Triggered)

Display Select 
(Edge Triggered)
Display Intensity Level 
(Edge Triggered)

Display Hold

Reset (Edge triggered)

Change backlight color with each
activation (backlight version only).

Advance once for each activation.

Increase intensity one level for each
activation (backlight version only).

Holds the assigned display, but all other
meter functions continue as long as
activated (maintained action).

Entering Code 50 will display the version (x.x) of the

meter. The display the returns to 

CodE 00

. Press the 

SEL

button to exit the module. 

ª

«

50

CodE

VIEW VERSION DISPLAY

Summary of Contents for DP63000A-I

Page 1: ...LP0694A omega com e mail info omega com For latest product manuals omegamanual info User sGuide Shop online at D DP P6 63 30 00 00 0A A I I D DP P6 63 30 00 00 0B B I I Digital P Panel M Meter...

Page 2: ...megashop cz Germany Austria Daimlerstrasse 26 D 75392 Deckenpfronn Germany TEL 49 0 7056 9398 0 FAX 49 0 7056 9398 29 Toll Free in Germany 0800 639 7678 e mail info omega de United Kingdom One Omega D...

Page 3: ...s Serial communications capability for RS232 or RS485 is added with a serial option module The DP63000 can be powered from an optional Power Supply Model Number DP6 MLPS1 that attaches directly to the...

Page 4: ...1 Recognized to U S and Canadian requirements under the Component Recognition Program of Underwriters Laboratories Inc UL Listed File E313547 UL508 CSA C22 2 No 14 M95 LISTED by Und Lab Inc to U S and...

Page 5: ...ver the rear of the unit to the back of the bezel The unit should be installed fully assembled Insert the unit into the panel cutout While holding the unit in place push the panel latch over the rear...

Page 6: ...commercial radio transmitter 4 Signal or Control cables within an enclosure should be routed as far as possible from contactors control relays transformers and other noisy components 5 In extremely h...

Page 7: ...tions must then be given to the potential of the user input common with respect to earth ground and the common of the plug in cards with respect to input common Before connecting signal wires the Inpu...

Page 8: ...he value and move to the next parameter PROGRAMMING MODE EXIT SEL BUTTON The Programming Mode is exited by pressing the SEL button with Pro NO displayed This will commit any stored parameter changes t...

Page 9: ...KEY style enter the first Input Value using the front panel buttons The Input Range selection sets the decimal location for the Input Value For Apply APLY style the meter shows the previously stored I...

Page 10: ...current calibration values to provide accurate current measurements Over time the electrical characteristics of the components inside the unit will slowly change with the result that the stored calib...

Page 11: ...n allows the SEL button to toggle through the enabled displays FRONT PANEL DISPLAY SELECT ENABLE SEL This selection allows the RST button to reset the selected value s FRONT PANEL RESET ENABLE RST The...

Page 12: ...en the setpoint is a control output usually balanced hysteresis is used For alarm applications usually unbalanced hysteresis is used For unbalanced hysteresis modes the hysteresis functions on the low...

Page 13: ...r input or RST button is activated momentary action the meter delays the event until the corresponding on output crosses the trigger off point Previously latched outputs are off if power up Display Va...

Page 14: ...PROGRAMMING QUICK OVERVIEW Press and hold SEL button to enter Programming Mode...

Page 15: ...therwise shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment sold...

Page 16: ...rs Turbine Paddlewheel Systems Totalizers Batch Controllers pH CONDUCTIVITY pH Electrodes Testers Accessories Benchtop Laboratory Meters Controllers Calibrators Simulators Pumps Industrial pH Conducti...

Reviews: