Appendix A
API Command Reference
NetScan User’s Manual
A-35
O
- Set Digital Outputs
TYPE
System
EXECUTION
Immediate
SYNTAX
Obank1,
bank2,
bank3,
bank4
Set any of the 32 digital outputs (separated into four 8-bit banks) to a
specified setting, where
bankn
(for
n
=
1
,
2
,
3
,
4
) is an argument in the
form of the number
nnn
such that
000
<
nnn
<
255
when converted to
binary format represents the desired settings for the 8-bit bank.
O?
Query current state of digital output.
DESCRIPTION
The Set Digital Outputs (
O
) command may be used to force any of the 32 digital outputs (separated into four 8-bit
banks) on the unit to a certain setting. The Bank to Digital Output mapping is as follows:
Bank
Digital Output
1
01 - 08
2
09 - 16
3
17 - 24
4
25 - 32
Each argument
bankn
represents the desired bit mapping for the corresponding bank of digital outputs. The
argument is a decimal number, which, when converted to binary format is the desired settings for that bank of
digital outputs. For example, an argument of
201
for
bank2
would be interpreted in binary as
11001001
(with
1
as active high and
0
as active low). This would affect the digital outputs of Bank 2 as follows:
Bank 2 Digital Outputs
9
10
11
12
13
14
15
16
high
high
low
low
high
low
low
high
Note: The setting of a bank will force the digital outputs to the specified setting regardless of the current state of
the digital outputs (such as if an alarm has been assigned to the output).
If it is desired to not affect a certain bank (maybe because it is being used for alarm outputs), a value of
999
can
be used for the argument for that bank. This will effectively “mask off” that bank from being updated.
There are 8 digital input lines and 32 digital output lines available on a rear panel DB50 connector. The digital
output can be controlled either “automatically” via the alarm settings or manually using the Set Digital Outputs (
O
)
command. To determine the digital output state, use the User Status (
U
) commands.
The Set Digital Outputs (
O
) command allows you to enter an output range and specify whether the output should
be cleared (logic false) or set (logic true). This command will force the specified digital output to the specified
condition. A
1
will drive the corresponding digital output(s) high. Conversely, a
0
will drive the corresponding
digital output(s) low.
Each digital output line will drive 5 standard TTL loads. All digital input lines are one-eighth (0.125) TTL loads. All
inputs are protected against damage from high static voltages. Normal precautions should be taken to limit the
input voltages to 0.3 to 5.3 volts. All digital I/O lines are referenced to digital ground.
EXAMPLE
PRINT#1, “O?X”
‘ Get the current state of the digital outputs banks
LINE INPUT #1, N$
PRINT N$
‘ Screen shows
O128,255,065,024
PRINT#1, “O0,999,76,234X”
‘ Set Banks 1, 3, and 4
PRINT#1, “O?X”
‘ Get the current state of the digital outputs
LINE INPUT #1, O$
PRINT O$
‘ Screen shows
O000,255,076,234
‘ Note that Bank 2 did not change
Summary of Contents for OMB-NETSCAN 1500
Page 6: ...iv NetScan User s Manual...
Page 18: ...1 12 Configuring and Starting NetScan NetScan User s Manual Notes...
Page 38: ...3 8 General Information and Specifications NetScan User s Manual Notes...
Page 82: ...4 44 ChartView Software Reference NetScan User s Manual Notes...
Page 118: ...6 20 Calibration NetScan User s Manual...
Page 140: ...A ii NetScan User s Manual...
Page 192: ...API Command Reference Appendix A A 52 NetScan User s Manual Notes...
Page 237: ...Appendix D Registers Data Formats Queries NetScan User s Manual D 13...
Page 244: ...NetScan Program Examples Appendix E E 2 NetScan User s Manual...
Page 248: ...ASCII Code Summary Appendix F F 4 NetScan User s Manual Notes...
Page 250: ...NetScan Error Messages Appendix G G 2 NetScan User s Manual Notes...
Page 252: ...Abbreviations Appendix H H 2 NetScan User s Manual Notes...
Page 254: ...NetScan User s Manual...