background image

2

2

The following describes the ordering information:

OS136 – * – ** , where

- * means:
-1 : -18 to 204ºC (0 to 400ºF) Temperature range
-2 :  149 to 538ºC (300 to 1000ºF) Temperature range

- ** means:
-MA : 4 to 20 mA current output
-V1 : 0 to 5 Vdc output
-V2 : 0 to 10 Vdc output
-K : K type thermocouple output
-MVC : 10 mV/ºC output
-MVF : 10 mV/ºF output

The following table lists the optional accessories: 

2.2 – Electrical Connection

The shielded cable provides the power and output connections. Fig 2-1 shows
the wiring diagram for different analog outputs.

Figure 2-1.  General Wiring Diagram

Accessories

Model No.

Description

OS100-MB

Mounting Bracket

OS136-WC

Water/Air Cooling Jacket

OS100-AP

Air Purge Collar

OS100-LS

Laser Sighting

PSU-93

Unregulated 16-24 VDC Power Supply

CAL-3-IR

NIST Traceable Calibration

Installation

RED

WHITE

YELLOW

THERMOCOUPLE

METER/

RECORDER

12-24 Vdc

POWER SUPPLY

GREEN

RED

-OR-

10 mV/Deg
0/5 Vdc or 4-20 mA
or 0/10 Vdc

EARTH
GROUND

DIGITAL

VOLTMETER/

AMMETER/

RECORDER

K TYPE 

THERMOCOUPLES

BLACK

SHIELD

+

+

+

OS136

WHITE

12-24 Vdc

POWER SUPPLY

EARTH

GROUND

BLACK

SHIELD

+

Summary of Contents for OS136 SERIES

Page 1: ...ost Non Contact Infrared Temperature Sensor Transmitter CAUTION This product is not intended for medical use or use on humans e mail info omega com For latest product manuals omegamanual info Shop onl...

Page 2: ...Managed by the United Kingdom Office Toll Free 0800 466 342 TEL 33 0 161 37 29 00 FAX 33 0 130 57 54 27 e mail sales omega fr Germany Austria Daimlerstrasse 26 D 75392 Deckenpfronn Germany Toll Free...

Page 3: ...mbols iii Section 1 Introduction 1 Section 2 Installation 1 2 1 Unpacking 1 2 2 Electrical Connection 2 Section 3 Operation 3 3 1 Measuring Temperature 3 3 2 Ambient Temperature 3 3 3 Atmospheric Qual...

Page 4: ...Description Page 2 1 General Wiring Diagram 2 3 1 Optical Field of View 3 3 2 Mounting Bracket OS100 MB 4 3 3 Air Purge Collar OS100 AP 4 3 4 Stainless Steel Housing 5 3 5 Water Air Cool Jacket OS136...

Page 5: ...The cable is a multi conductor 24 AWG stranded wire with the rating of 600 VDC 205 C 401 F Teflon insulation Power must be disconnected before making any electrical connections The recommended power...

Page 6: ...4 rated The unit comes with a 1 82 m 6 shielded cable as standard SECTION 2 INSTALLATION 2 1 Unpacking Remove the packing list and verify that you have received all your equipment If you have any que...

Page 7: ...r and output connections Fig 2 1 shows the wiring diagram for different analog outputs Figure 2 1 General Wiring Diagram Accessories Model No Description OS100 MB Mounting Bracket OS136 WC Water Air C...

Page 8: ...tter can operate in an ambient temperature of 0 to 70 C 32 to 158 F without any water cool jacket It can operate from 0 to 200 C 32 to 392 F with the water cool jacket accessory OS136 WC Fig 3 5 3 6 I...

Page 9: ...e following figures show the Mounting Bracket OS136 MB Air Purge Collar OS136 AP Stainless Steel Housing Water Air Cool Jacket OS136 WC and the Water Air Cool Jacket Assembly Figure 3 2 Mounting Brack...

Page 10: ...HEX KNURL 4 76 3 16 WIDE DIMENSIONS mm in 3 4 16 UNF 88 9 3 5 1 83m 6ft 33 27 1 31 19 05 0 75 1 8 NPT COMPRESSION FITTING 2 PLCS 34 80 1 37 85 09 3 35 7 62 0 30 3 4 16 UNF 2B 3 4 16 UNF INLET OUTLET D...

Page 11: ...de 4 2 Operating the Laser Sight Accessory The Laser sight accessory screws onto the front of the transmitter sensor head This accessory is only used for alignment of the transmitter head to the targe...

Page 12: ...5 POWER CABLE ON OFF SWITCH BATTERY PACK POWER SUPPLY OS136 Dimensions mm in OUTPUT 1 mW WAVELENGTH 630 670 nm CLASS II 2 LASER PRODUCT COMPLIES WITH FDA 21CFR 1040 10 EN60825 1 11 2001 LASER RADIATIO...

Page 13: ...uirements Min Load 0 to 5 Vdc 1 K Ohms Min Load 0 to 10 Vdc 2 K Ohms Max Load 4 to 20 mA Power Supply 4 20 mA Min Load 10 mV Deg 10 K Ohms Min Load K T C 100 K Ohms Operating Ambient Temperature No Wa...

Page 14: ...er Beam Diameter Less than 5 mm Beam Divergence Less than 2 mrad Operating Temperature 0 to 50 C 32 to 122 F Operating Relative Humidity Less than 95 RH non condensing Power Switch ON OFF Slide switch...

Page 15: ...therwise shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment sold...

Page 16: ...ity Indicators Turbine Paddlewheel Systems Totalizers Batch Controllers pH CONDUCTIVITY pH Electrodes Testers Accessories Benchtop Laboratory Meters Controllers Calibrators Simulators Pumps Industrial...

Reviews: