19. Press the SENSOR SELECT function button. The sensor type will
change to type T. The “J” will disappear from the display and a “T” will
appear on the display. The voltmeter should read 0.000 mV.
20. Press the SENSOR SELECT function button. The sensor type will
change to type E. The “T” will disappear from the display and an “E” will
appear on the display. The voltmeter should read 0.000 mV.
21. Press the SENSOR SELECT function button. The sensor type will
change to type N. The “E” will disappear from the display and a “N” will
appear on the display. The voltmeter should read 0.000 mV.
22. Press the SENSOR SELECT function button. The sensor type will
change to type B. The “N” will disappear from the display and a “B” will
appear on the display. “LO” will also appear on the display.
23. Press the CHANGE function button. The “32.0 °F” display will blink, and
“NUM” will appear on the display.
24. Press 1 0 0 0 0 buttons and then press ENTER function button. “NUM”
will disappear and “1000.0 °F” will appear, not blinking. The voltmeter
should read 1.438 mV.
25. Press the SENSOR SELECT function button. The sensor type will
change to type R. The “B” will disappear from the display and a “R” will
appear on the display.
26. Press the CHANGE function button. The “1000.0 °F” display will blink,
and “NUM” will appear on the display.
27. Press 3 2 0 buttons and then press ENTER function button. “NUM” will
disappear and “32.0 °F” will appear, not blinking. The voltmeter should
read 0.000 mV.
28. Press the SENSOR SELECT function button. The sensor type will
change to type S. The “R” will disappear from the display and a “S” will
appear on the display. The voltmeter should read 0.000 mV.
29. Press the SENSOR SELECT function button. The sensor type will change to
type G. The “R” will disappear from the display and a “g” will appear momen-
tarily in the center of the display. “LO” will then appear in the center of the dis-
play and both the “T” and “R” will appear along the top row of the display.
30. Press the CHANGE function button. The “32.0 °F” display will blink, and
“NUM” will appear on the display.
31. Press 6 0 0 0 buttons and then press ENTER function button. “NUM” will dis-
appear and “600.0 °F” will appear, not blinking. The voltmeter should read
2.238 mV.
35