24
ENTERING OR CHANGING YOUR (DEFAULT) ID CODE:
Enter
ID
menu (Repeat steps from 1 to 3).
Press
d
10)
Display advances to
CH.ID
Change ID Code Submenu.
Press
d
11)
Display shows
0000
message with flashing 1
st
digit.
If you want to change your default “ID Code” you can do it now, otherwise
press
a
and menu will skip to
FULL
Full Security Submenu.
Press
b
&
c
12)
Press
b
and
c
buttons to enter your new “ID Code” number.
Press
d
13)
Display shows
STRD
stored message momentarily and then
advances to the
FULL
Full Security Submenu.
FULL SECURITY LEVEL SUBMENU:
Press
d
14)
Display flashes
ENBL
Enable or
DSBL
Disable.
Press
b
15)
Scroll through the available selections: “Enable” or “Disable”.
Press
d
16)
Display shows
STRD
stored message momentarily and then
advances to
SP.ID
Setpoint/ID Submenu.
If "Full" Security Level is "Enabled" and the user attempts to enter the
Main Menu, they will be prompted for an ID Code. The ID Code should
be correct to enter the instrument Menu item.
SETPOINT/ID SECURITY LEVEL SUBMENU:
This Security Level can be functional only if
FULL
Security Level
is Disabled.
Press
d
17)
Display flashes
ENBL
Enable or
DSBL
Disable.
Press
b
18)
Scroll through the available selections: “Enable” or “Disable”.
Press
d
19)
Display shows
STRD
stored message momentarily and then
advances to
COMM
Communication Submenu.
If
"Setpoint/ID"
Security Level is
"Enabled"
and the user attempts to advance
into the
CNFG
Configuration Menu, he will be prompted for ID Code number.
The ID Code should be correct to proceed into the Configuration Menu,
otherwise display will show an Error and skip to the Run Mode.
If
“Full”
and
“Setpoint/ID”
Security Levels are "
Disabled
", the ID code will be
“Disabled” and user will not be asked for ID Code to enter the Menu items (“ID”
Submenu will not show up in “ID/Setpoint” Menu).