The battery icon on the LCD shows the status of the batteries as shown below.
M4098-BatteryIcons.eps
Full Battery Life (100%)
75% Battery Life
50% Battery Life
25% Battery Life
Low Battery - Replace Battery
No Battery Icon – AC adapter
Figure 2-6. Battery Warnings
2.3.2 Connecting ac Power
The recorder can be ac powered using the universal 100-240 Vac adapter
supplied. The dc power jack is located on the right side of the unit (refer to
Figure 1-1, item #6.). The ac adapter provides 9 Vdc @ 1.7A power output and
comes with a 1.8 meters (6 ft.) long cable.
The following parameters are stored in the non-volatile
memory and will not be lost when batteries or main power
are removed
• Chart speed & scale
• High & low alarm set points for the two channel
• Chart Data points
When the recorder runs only on AC adaptor, and the main power restores
after a power failure, the recorder turns on automatically without pressing the
Power Key. All LCD segments turn on momentarily, then there is a 20 second
delay before the revision screen. This only happens when the power is removed
without pressing the Power Key. The pens go to the home position and back to
the correct chart scale everytime the recorder powers up.