In order to calibrate the input channels, you need to enter the password “caldtc”.
Then you can calibrate each input channel. At every calibration point, you need
to simulate the thermocouple input for the calibration point temperature, then
press the ‘Set’ key to enter calibration data. At the end of your calibration, press
“Save’ to complete the calibration process. Here are the calibration points for the
thermocouple input types:
Low Calibration Point High Calibration Point
K type
100°F
1100°F
J type
100°F
500°F
T type
100°F
500°F
Figure 3-12. Settings Menu (Re-scale/Calibrate/Memory Tab) Screen