Model DFG-RS3 Digital Force/Torque Indicator
User’s Guide
4
Example 1
Model RTQ50-50Z sensor with Model DFG-RS3 Indicator
RTQ50-50Z
±0.35% of full scale
+
DFG-RS3
±0.2% of full scale
=
Total
±0.55% of full scale
This translates into a fixed error of up to:
0.55% x 50 ozFin = 0.275 ozFin
Example 2
Model MR01-100 sensor with Model DFG-RS5 Indicator
RLC01-100
±0.15% of full scale
+
DFG-RS5
±0.1% of full scale
=
Total
±0.25% of full scale
This translates into a fixed error of up to:
0.25% x 100 lbF = 0.25 lbF
Because accuracy is defined as a
percentage of full scale
, the fixed error is possible anywhere on the
scale from 0 to the capacity. As such, this value represents an increasingly large error as
percentage of
reading
towards the low end of the scale. It is, therefore, recommended that a sensor is selected with
capacity as close as possible to the expected load.
The resolution may be different for some sensors depending on whether the DFG-RS5 or DFG-RS3
indicator is used. For example, a Series RLC01 force sensor will display finer resolution when connected
to a DFG-RS5 indicator than when connected to a DFG-RS3 indicator. Resolution information is shown in
the sensors’ user’s guide.
1.4 Safety / Proper Usage
Read through the following safety instructions thoroughly before using the DFG-RS3 with a sensor:
1. Note the sensor’s capacity before use and ensure that the capacity is not exceeded.
Producing a
load greater than the indicated safe overload value can damage the sensor.
An overload
can occur whether the sensor’s indicator is powered on or off.
2. In order to extend the life of the sensor, avoid repetitive shock and impact loading.
3. When moving the sensor to another location, never lift from the cable or strain relief. This can
cause damage to the sensor. Always lift the sensor housing itself.
4. Always ensure that load is applied axially with respect to the sensor.
5. Ensure that the sensor is kept away from water or any other electrically conductive liquids at all
times.
6. The sensor and indicator should be serviced by a trained technician only. AC power must be
disconnected and the indicator must be powered off before the housing is opened.
7. Always consider the characteristics of the sample being tested before initiating a test. A risk
assessment should be carried out beforehand to ensure that all safety measures have been
addressed and implemented.
Summary of Contents for DFG-RS3
Page 1: ...Model DFG RS3 Digital Force Torque Indicator...
Page 2: ...Model DFG RS3 Digital Force Torque Indicator User s Guide 1...
Page 22: ...Model DFG RS3 Digital Force Torque Indicator User s Guide 21 12 3 Dimensions IN MM...
Page 23: ...Model DFG RS3 Digital Force Torque Indicator User s Guide 22...
Page 24: ...Model DFG RS3 Digital Force Torque Indicator User s Guide 23 M 5250 1017...