Model DFG-RS3 Digital Force/Torque Indicator 

 

User’s Guide 

 
 

 

 

 

4

 

Example 1

 

Model RTQ50-50Z sensor with Model DFG-RS3 Indicator 

 

RTQ50-50Z 

±0.35% of full scale

 

 + 

DFG-RS3 

±0.2% of full scale 

 

 = 

Total 

±0.55% of full scale 

 

 

This translates into a fixed error of up to: 
0.55% x 50 ozFin = 0.275 ozFin  

 
Example 2 

Model MR01-100 sensor with Model DFG-RS5 Indicator 

 

RLC01-100 

±0.15% of full scale

 

 + 

DFG-RS5 

±0.1% of full scale 

 

 = 

Total 

±0.25% of full scale 

 

 

This translates into a fixed error of up to: 
0.25% x 100 lbF = 0.25 lbF  

 

Because accuracy is defined as a 

percentage of full scale

, the fixed error is possible anywhere on the 

scale from 0 to the capacity. As such, this value represents an increasingly large error as 

percentage of 

reading

 towards the low end of the scale. It is, therefore, recommended that a sensor is selected with 

capacity as close as possible to the expected load. 
 
The resolution may be different for some sensors depending on whether the DFG-RS5 or DFG-RS3 
indicator is used. For example, a Series RLC01 force sensor will display finer resolution when connected 
to a DFG-RS5 indicator than when connected to a DFG-RS3 indicator. Resolution information is shown in 
the sensors’ user’s guide. 

 

 

1.4 Safety / Proper Usage 

 
Read through the following safety instructions thoroughly before using the DFG-RS3 with a sensor: 
 

1.  Note the sensor’s capacity before use and ensure that the capacity is not exceeded. 

Producing a 

load greater than the indicated safe overload value can damage the sensor.

 An overload 

can occur whether the sensor’s indicator is powered on or off. 

 

2.  In order to extend the life of the sensor, avoid repetitive shock and impact loading. 

 

3.  When moving the sensor to another location, never lift from the cable or strain relief. This can 

cause damage to the sensor. Always lift the sensor housing itself.  

 

4.  Always ensure that load is applied axially with respect to the sensor.  

 

5.  Ensure that the sensor is kept away from water or any other electrically conductive liquids at all 

times. 

 

6.  The sensor and indicator should be serviced by a trained technician only. AC power must be 

disconnected and the indicator must be powered off before the housing is opened. 

 

7.  Always consider the characteristics of the sample being tested before initiating a test. A risk 

assessment should be carried out beforehand to ensure that all safety measures have been 
addressed and implemented.  

 

Summary of Contents for DFG-RS3

Page 1: ...Model DFG RS3 Digital Force Torque Indicator...

Page 2: ...Model DFG RS3 Digital Force Torque Indicator User s Guide 1...

Page 3: ...provides setup safety and operation instructions Dimensions and specifications are also provided For additional information or answers to your questions please do not hesitate to contact us Our techn...

Page 4: ...indicator and connected to another without the need for re calibration or re configuration All such data is saved within a PCB located inside the smart connector The model number serial number and ca...

Page 5: ...connected to a DFG RS3 indicator Resolution information is shown in the sensors user s guide 1 4 Safety Proper Usage Read through the following safety instructions thoroughly before using the DFG RS3...

Page 6: ...f suitably rugged materials and components Similar precautions should be taken when mounting the sensor to a test stand work bench or other piece of equipment 2 POWER The DFG RS3 is powered either by...

Page 7: ...into place with a click Fig 3 1 Appropriate orientation of the connector To release the connector press both buttons on either side of the indicator housing to release the sensor as shown in Fig 3 2 b...

Page 8: ...Primary reading The current displayed load reading See Operating Modes section for details If a sensor is not plugged in this value will be replaced by a message as follows SENSOR NOT CONNECTED 4 Load...

Page 9: ...upper load limit the displayed value is between the load limits the displayed value is less than the lower load limit 9 Set points The programmed load limit values Typically used for pass fail type te...

Page 10: ...e As described above the Plug TestTM smart connector retains all configuration and calibration data for the sensor which includes menu settings As such a sensor must be connected in order for menu cha...

Page 11: ...corner of the display will be blank When set points are enabled the following indicators are shown to the left of the primary reading the displayed value is greater than the upper load limit NO GO HIG...

Page 12: ...selected 9 COMMUNICATIONS AND OUTPUTS Communication with the DFG RS3 is achieved through the micro USB port located along the left side of the housing as shown in the illustration in the Power section...

Page 13: ...ssion clockwise values have positive polarity tension counter clockwise values have negative polarity Numeric Only Output format includes the value only Polarity same as above Invert Polarity Compress...

Page 14: ...nsor can be calibrated at up to 10 points in each direction Enter the number of calibration points for each direction compression and tension or clockwise and counter clockwise At least one point must...

Page 15: ...quired Do not yet attach any weights or apply any calibration loads Press ENTER 6 The display appears as follows Optionally exercise the sensor several times at full scale if possible then press ENTER...

Page 16: ...F capacity sensor is calibrated and 5 data points were selected the load values will default to 10 20 30 40 and 50 lb Apply the calibration load Then press ENTER Repeat the above step for the number o...

Page 17: ...pter ensure that the output signal leads SG and SG have been installed into the appropriate terminal blocks Some sensor manufacturers consider SG to be a compression value while others consider it to...

Page 18: ...sed 11 2 Backlight There are several available initial settings applicable upon powering on the indicator To access these settings select Backlight from the menu The display appears as follows Selecti...

Page 19: ...functions for which audible tones will apply select Tones from the menu The screen appears as follows 11 5 Initial Mode This section is used to configure the initial mode upon powering on the indicat...

Page 20: ...nformation Welcome Screen The following screen is displayed at power up and can be accessed at any time by selecting Information from the menu RESTORE DEFAULT SETTINGS No Yes Digital Indicator Model D...

Page 21: ...0 of full scale display shows OVER at 110 and above Weight 0 7 lb 0 3 kg Included accessories Carrying case AC adapter battery USB cable resource CD USB driver and user s guide certificate of conforma...

Page 22: ...Model DFG RS3 Digital Force Torque Indicator User s Guide 21 12 3 Dimensions IN MM...

Page 23: ...Model DFG RS3 Digital Force Torque Indicator User s Guide 22...

Page 24: ...Model DFG RS3 Digital Force Torque Indicator User s Guide 23 M 5250 1017...

Reviews: