22
TABLE 4-1
CALIBRATION VALUES WHEN USING
THERMOCOUPLE CALIBRATOR
Look at the end of Table 4-1 for the explanation of .
* Look at Table 1-1, for complete models (preface section).
MODEL
T/C CALIBRATOR METER
POT TO
OPTION
DISPLAY
DISPLAY
ADJUST
*-JF1
32¡F
32¡F
Zero
1346¡F
1346¡F
Span
*-JF2
32.0¡F
31.6¡F
Zero
167.0¡F
167.0¡F
Span
*-JC1
0¡C
0¡C
Zero
730¡C
730¡C
Span
*-JC2 0.0¡C
-0.2¡C
Zero
170.0¡C
170.0¡C
Span
*-KF1
32¡F
31¡F
Zero
1794¡F
1794¡F
Span
*-KF2
32.0¡F
32.0¡F
Zero
160.0¡F
160.0¡F
Span
*-KC1
0¡C
0¡C
Zero
1335¡C
1335¡C Span
*-KC2
0.0¡C
0.0¡C
Zero
190.0¡C
190.0¡C Span