23
TABLE 4-1 (Cont’d)
The best MID-RANGE accuracy is obtained when the
meter is calibrated to the exact values shown in column 3,
even though a small error may occur at 32¡F (0¡C).
*
Look at Table 1-1, for complete models (preface section).
MODEL
T/C CALIBRATOR METER
POT TO
OPTION
DISPLAY
ADJUST
DISPLAY
*-TF1
32¡F
31¡F
Zero
716¡F
716¡F Span
*-TF2
32.0¡F
32.0¡F
Zero
185.0¡F
185.0¡F Span
*-TC1
0¡C
0¡C
Zero
380¡C
380¡C Span
*-TC2
0.0¡C
0.0¡C
Zero
185.0¡C
185.0¡C Span
*-EF1
32¡F
31¡F
Zero
1769¡F
1769¡F Span
*-EF2
32.0¡F
33.0¡F
Zero
180.0¡F
180.0¡F
Span
*-EC1
0¡C
0¡C
Zero
965¡C
965¡C
Span
*-EC2
0¡C
0.5¡C
Zero
185.0¡C
185.0¡C
Span