26
TABLE 4-2
CALIBRATION VALUES
WHEN USING A MILLIVOLT SOURCE
Look at the end of Table 4-2, for the explanation of .
* Look at Table 1-1, for complete models (preface section).
MILLIVOLT
MODEL
VALUE FROM
METER
POT TO
OPTION
MILLIVOLT DISPLAY
ADJUST
SOURCE
*-JF1
0.0
32¡F
Zero
41.013
1346¡F
Span
*-JF2
0.0
31.6¡F
Zero
3.917
167.0¡F
Span
*-JC1
0.0
0¡C
Zero
41.013
730¡C
Span
*-JC2
0.0
-0.2¡C
Zero
9.113
170.0¡C
Span
*-KF1
0.0
31¡F
Zero
40.449
1794¡F
Span
*-KF2
0.0
32.0¡F
Zero
2.896
160.0¡F
Span
*-KC1
0.0
0¡C
Zero
53.611
1335¡C
Span
*-KC2
0.0
0.0¡C
Zero
7.737
190.0¡C
Span