27
TABLE 4-2 (Cont’d)
The best MID-RANGE accuracy is obtained when the
meter is calibrated to the exact values shown in column 3,
even though a small error may occur at 32¡F (0¡C).
* Look at Table 1-1, for complete models (preface section).
MILLIVOLT
MODEL
VALUE FROM
METER
POT TO
OPTION
MILLIVOLT
DISPLAY
ADJUST
SOURCE
*-TF1
0.0
31¡F
Zero
19.638
716¡F
Span
*-TF2
0.0
32.0¡F
Zero
3.584
185.0¡F
Span
*-TC1
0.0
0¡C
Zero
19.638
380¡C
Span
*-TC2
0.0
0.0¡C
Zero
8.495
185.0¡C
Span
*-EF1
0.0
31¡F
Zero
73.727
1769¡F
Span
*-EF2
0.0
33.0¡F
Zero
5.130
180.0¡F
Span
*-EC1
0.0
0¡C
Zero
73.727
965¡C
Span
*-EC2
0.0
0.5¡C
Zero
12.314
185.0¡C
Span