START HERE

2

3

4

Thermocouple Wire Connection

1.

Connect positive (+) lead of thermocouple.

2.

Connect negative (-) lead of thermocouple.

Note: The negative lead is red.

Example hook up for AC Load

Alarm 1 (Setpoint) Hook-up

1.

Connect a jumper from ac Line to Relay 1 Common
(Terminal 3).

2.

Connect Relay 1 Normally Open (Terminal 1) to External
Alarm ac Line.

3.

Connect External Alarm to ac Neutral.

Alarm 2 (Setpoint) Hook up

1.

Connect a jumper from ac Line to Relay 2 Common
(Terminal 6).

2.

Connect Relay 2 Normally Open (Terminal 4) to External
Alarm ac Line.

3.

Connect External Alarm to ac Neutral.

Analog Output Wiring for 4 - 20 mA Current

1.

Connect Positive Lead to Terminal 11.

2.

Connect Negative Lead to Terminal 12.

For 0 -10 Voltage

1.

Connect Positive Lead to Terminal 10.

2.

Connect Negative Lead to Terminal 12.

Using This Quick Start Manual

Use this Quick Start Manual with your controller to make
changes to the thermocouple type, decimal point, units, and to
change the setpoints.

Features with        are for the “B” version which has three-
color programmable “Big” LED display - All segment
characters shown are for the “B” version.

For detailed instructions, refer to the appropriate section
in the Operator’s Manual.

Before You Begin

In addition to the meter and the related parts, you will need
the following items to set up your meter:

ac power, as listed on meter’s ID/Power Label

Thermocouple

1/8” flat blade screwdriver

Mount the Unit

1. Cut a panel opening using

the dimensions shown to
the right.

2. Position the unit in the

opening, making sure the
front bezel is flush with the
panel.

3. Install retaining clip on the

meter and tighten against
the panel.

Wiring

1. Remove the panel at the back of the unit.
2. Locate the TB1 connector.
3. Insert the correct wire in each terminal as shown in the

following figure and tighten the lockdown screws.

4. Tug gently on the wires to verify the connections.

AC Powered Unit Connections

External Fuse Required:

Time-delay, UL 248-14 listed

Time-lag, IEC 127-3 recognized

175 mA (115 Vac line)

125 mA (115 Vac line)

80 mA (230 Vac line)

63 mA (230 Vac line)

In order to maintain the same degree of
protection as the AC units, always use a
Safety Agency Approval DC source with
the same Overvoltage Category and
Pollution Degree.

-

+

TB4B

TB4C

-

+

TB4B

TB4C

Warning: Do not connect AC power to your device until you
have completed all input and output connections. This
device must only be installed by a specially trained
electrician with corresponding qualifications. Failure to
follow all instructions and warnings may result in injury!

Safety Consideration

This device is marked with the international Caution symbol.

The instrument is a panel mount device protected in
accordance with EN61010-1 (Safety requirements for electrical
equipment for measurement, control, and laboratory standard).
Remember that the unit has no power-on switch. Building
installation should include a switch or circuit-breaker that must
be compliant to IEC 947-1 and 947-3.

SAFETY:

• Do not exceed voltage rating on the label located on the

top of the instrument housing.

• Always disconnect power before changing signal and

power connections.

• Do not use this instrument on a work bench without its

case for safety reasons.

• Do not operate this instrument in flammable or explosive

atmospheres.

• Do not expose this instrument to rain or moisture.

EMC:

• Whenever EMC is an issue, always use shielded cables.
• Never run signal and power wires in the same conduit.
• Use signal wire connections with twisted-pair cables.
• Install Ferrite Bead(s) on signal wire close to the instrument

if EMC problems persist.

Wiring the Controller

Typical Wiring of TB1

Using the Configuration Menu

To configure the meter, you use the buttons on the front panel.

To:

Take This Action:

Display the

Press the 

MENU

button. The first function

Configuration Menu on the menu, 

INPT

, displays.

Select a submenu

1. Press 

MENU

until the function you

function

want is shown.

2. Press 

/DEV

.

The information you can change flashes.

Select a value

1. Press 

/MAX

to display the option

for that submenu

you want.

function

2. Press 

MENU

to store it.

STRD

quickly flashes, indicating that

the selection has been stored in memory. 
Then the next menu function displays.

Go back to previous Press 

RESET

once.

menu function
Exit the

Press 

RESET

twice. The unit displays 

Configuration

RST

as it reinitializes. When a numeric

Menu

value displays, the unit is in run mode.
(Optionally, you can press 

MENU

to

move through all the menu functions
until the unit reinitializes.)

MENU 

SUBMENU      

/DEV

DESCRIPTION

INPT

J.TC

,

K.TC

*,

DJ.TC

T.TC 

Input

DEC.P

FFFF

*,

FFF.F

Decimal Point

RD.CF

R.1=C

,

R.1=F

*

Reading Configuration

COLR

GRN

,

RED

,   

AMBR

Display Color

S1.CF

S.1=A

*,

S.1=B

Setpoint 1 Configuration

S.2=U

*,

S.2=L

S2.CF

S.1=A

*,

S.1=B

Setpoint 2 Configuration

S.2=U

*,

S.2=L

S1.DB

0003

*

Setpoint 1, Deadband

S2.DB

0003

*

Setpoint 2, Deadband

OT.CF

O.1=E

*, 

0.1=D

Analog Output

O.2=C

*, 

0.2=V

Configuration

O.3=A

*, 

0.3=P

O.4=D

,  

0.4=R

O.5=F

,  

0.5=H

P.BND

0000 

shown if 0.3 = P

Proportional Band

M.RST

0000 

shown if 0.3 = P

Manual Reset

OT.S.O

RD!1

,

OUT1

,

RD!2

,

OUT2

Output Scale & Offset

CJ.OF

0000

Cold Junction Offset

LK.CF

RS=E

*,

RS=D

Lockout Configuration

SP=E

*,

SP=D

L3=0

*,

L3=1

BRIT

M.BrT

,

L.BrT

,

H.BrT

Display Brightness

* Factory Default Settings.

DC Powered Unit Connections

Using the Menus

To Change the Thermocouple Type:

1.

Press MENU until the display
shows 

INPT

2.

Press  

/DEV  to show current

thermocouple type:

3.

Press  

/MAX  to select the

setting from J, K, Tor DJ.TC.

4.

Press MENU to store the value.

5.

Press RESETtwice to display the current temperature.

SETPTS

/DEV

MENU

RESET

/MAX

SETPTS

/DEV

MENU

RESET

/MAX

SETPTS

/DEV

MENU

RESET

/MAX

SETPTS

/DEV

MENU

RESET

/MAX

SETPTS

/DEV

MENU

RESET

/MAX

Reviews: