43
Configuring The Meter
4
43
4.14 USING LOCK OUT CONFIGURATION
LK.CF
Refer to Table 6-1 for a summary list of menu configuration.
Use Lock Out Configuration
LK.CF
for the following:
•
To enable or disable setpoint changes
•
To enable or disable the
RESET
button in the Run Mode
•
To enable or disable displaying the meter’s firmware version.
4.14.1 Enabling or Disabling the RESET button in the Run Mode
1.
Press the
MENU
button until the meter shows
LK.CF
(after
OT.S.O
).
2.
Press the
∂
TARE
button. The meter shows one of the following:
•
RS.=E
To enable the RESET button in the Run Mode
(Default)
•
RS.=D
To disable the
RESET
button in the Run Mode
3.
Press the
ß
MAX
button to toggle between available choices.
4.
Press the
MENU
button to store the changes. The meter shows
STRD
if the new
value is different otherwise the meter shows
BRIT
and returns to the Run Mode.
4.14.2 Enabling or Disabling SETPOINT Changes
1.
Press the
MENU
button until the meter shows
LK.CF
(after
OT.S.O
).
2.
Press the
∂
TARE
button twice. The meter shows one of the following:
•
SP.=E
To enable setpoint changes
(Default)
•
SP.=D
To disable setpoint changes
3.
Press the
ß
MAX
button to toggle between available choices.
4.
Press the
MENU
button to store the changes. The meter shows
STRD
if the new
value is different otherwise the meter shows
BRIT
and returns to the Run Mode.
Note
☞
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com