DP6060 Dual-Input Process Meter
Instruction Manual
98
Display Parameter Action/Setting
Description
Round
Round
Enter the round menu
RSt 1
Reset 1
Program reset point 1 (1-8)
Rst Hi
Reset high
Press Enter to reset max display
Rst HL
Reset high &
low
Press Enter to reset max & min displays
Rst Lo
Reset low
Press Enter to reset min display
Rst tr
Reset tare
Reset tare
Sampl
Sampling
Set relay for sampling operation
SCAL A
Scale A
Enter the
Scale
menu for channel A
SCAL B
Scale B
Enter the
Scale
menu for channel B
SelEct
Select
Enter Select menu (function, round, math,
constant, cutoff, output programming)
send
Send
Send meter settings to another meter
seriAl
Serial
Set serial communication parameters
Set 1
Set 1
Program set point 1 (1-8)
setup
Setup
Enter
Setup
menu
SlavId
Slave ID
Set Slave ID or meter address
Source
Source
Select source for the 4-20 mA output
Square
Square root
Set meter for square root extraction
Sunm
Sum
Channel C = (A+B+P)*F
Tare A
Tare A
Activate tare on channel A
Tare b
Tare B
Activate tare on channel B
Tr dLY
Transmit delay
Set transmit delay for serial communication
units
units
Enter in channel A, B, and C input
units/tags
unloc
Unlocked
Program password to lock meter
up
Up arrow
Assign digital input to Up arrow button/F2
u-rang
Underrange
Program mA output for display underrange
user
User I/O
Assign function keys and digital I/O
v CAL
Voltage
calibration
Calibrate voltage input
v Hi
Voltage high
Calibrate high voltage input (e.g. 10 V)
v Lo
Voltage low
Calibrate low voltage input (e.g. 0 V)
Volt
Volt
Set meter for volt input
uwAvg
Weighted avg.
Channel C = ((B-A)*F)+A
Troubleshooting Tips
Symptom Check/Action
No display at all
Check power at power connector
Not able to change setup or
programming,
Locd
is displayed
Meter is password-protected, enter
correct six-digit password to unlock
Meter displays error message
during calibration (
Error
)
Check:
1. Signal
connections