RTD THERMOMETER
6
9. TEMPERATURE METER CALIBRATION SETUP
Below table diagram shows each button function when user enters
into the calibration mode.
Note:
Turn the Power “OFF” before attempting SETUP.
Setup mode is cancelled during below procedure if “POWER”
button is pressed.
9.1 Temperature Calibration
Place the temperature probe in a known, stable temperature environment. Allow the readings to
stabilize. In Setup, change the offset until the display reading matches the calibration temperature.
Press and hold
+
buttons and then power on the meter to enter setup mode.
Press
button to enter calibration mode. (in 3 seconds)
Press
button to enter calibration mode. (in 3 seconds)
Press
or
button to confirm present Temperature value.
Press
button twice to confirm selection.
Note:
To abort during the process, press
button to exit calibration mode.
9.2 Recall Default factory setting value
Press and hold
+
buttons and then power on the meter to enter setup mode.
Press
button to enter calibration mode. (in 3 seconds)
Press
button to enter recall mode. (in 3 seconds)