To prevent accidental loss of MIN, MAX and AVG data, this 
mode  can  only  be  cancelled  by  pressing  and  holding  the 
“MIN/MAX”  key  for  2  seconds.  All  recorded  readings  are 
erased. 
 

7. REL Relative mode (only Main display)

 

Press  the  “REL”  key  to  enter  the  relative  mode,  zero  the 
display, and store the displayed Reading as a reference value. 
REL is shown on the

 

display. Press “REL” key again to exit 

the  relative  mode.  The  relative  reference  value  can  also  be 
entered by  the user. (See  “SET  mode” later in  this  manual). 
When  the  desired  relative  value  has  been  entered,  press 
“REL”  key  to  enter the  relative mode  and than press  “SET” 
key use the entered relative value as a reference value. Press 
“REL”  key  again  to  exit  the  relative  mode.  In  the  relative 
mode,  the  value  (can  not  >±1999.9  counts)  shown  on  the 
LCD  is  always  the  difference  between  the  stored  reference 
and the present reading. 
 

8. HOLD mode (only Main display) 

Press  the  “HOLD”  key  to  enter  the  Data  Hold  mode,  the 
“HOLD”  annunciator  is  displayed.  When  HOLD  mode  is 
selected, the thermometer held the present readings and stops 
all  further  measurements.  Press  the  “HOLD”  key  again  to 
cancel  HOLD  mode  causing  thermometer  to  resume  taking 
measurements.  In  the  MIN/MAX  recording  mode,  press 
“HOLD” key to stop the recording. Press “HOLD” key again 
to  resume  recording.  (Previously  recorded  read  are  not 
erased). 
 

9. SET mode (Relative value set, Time set and 

Hi/Lo Limits value set) 

9.1  Press  the  “SET”  key  to  enter  relative  values  SET  mode 

(Press “ENTER” key  to  skip  setting  relative  value).  =  = 
= =.= is displayed on the main display. The relative value 
is  entered  via  overlay  numbers,  press  overlay  “ENTER” 
key  to  store  the  relative  value,  and  advance  to  elapsed 
time set mode. 

9.2  Elapsed  time  set  mode,  (Press  ENTER  key  to  skip 

Elapsed time set mode) =.= = = : = = is displayed in sec-
ond  and  third  display.  Time  (hours,  minutes,  seconds) 
value  is  entered  via  overlay  numbers,  press  overlay 
“ENTER”  key  to  store  time  value.  Elapsed  time  starts 
from set time value. 

9.3 Hi Limit value set mode, (Press “ENTER” key to skip Hi 

Limit  value  set  mode),  =  =  =  =.=  is  displayed  in  main 
display,  Hi  Limit  value  is  entered  via  overlay  numbers, 
then press “ENTER” key to store the Hi Limit value. = = 
= =.= is displayed in main display, Lo Limit value is en-
tered  via  overlay  numbers, then  press  overlay  “ENTER” 
key to store the Lo Limit value and exit SET mode. 

9.4  When  the  thermometer  is  turned  on  it  uses  the  relative 

value  and  Hi/Lo  Limits  values  that  were  entered  when 
thermometer was last in use. 

 

10. T1/T2 T1-T2 second display Input Selection 

The  input  selection  indicates  which  input  is  selected  for 
second display; T1 probe, T2 probe or the difference between 
the two probes (T1-T2), when the thermometer is turned on, 
it  is  set  to  T2  input  can  select  alternate  of  probe  by  second 
display Pt385/Pt3926/Pt3916 key switch. 
 
11.  TYPE(Pt385/Pt3926/Pt3916)  Input  RTD  Probe 

select (second display) 
The  Pt385/Pt3926  key  switches  the  T2  input  to  select  the 
Pt385,  Pt3926  or  Pt3916  RTD  probe  as  input,  when  second 
display input selected T2. When the thermometer is turned on, 
it  is  set  to  the  RTD  selected  that  was  in  use  when  the  ther-
mometer was last turned off. 
 

12. Hi/Lo Limits mode (only Main display)

 

Press “Hi/Lo Limits” key to enter the Hi/Lo Limits compara-
tive mode. When the input temperature value exceeds the Hi 
or  Lo  Limits  value, the beeper  emits  a continuous  or pulsed 
tone. Press “Hi/Lo Limits” key again to exit the Hi/Lo Limits 
mode. 
 

RTD PROBE CONNECTION 

P/N: 14061-0800-9    MINI DIN PLUG 

Rear view 

 

 

Front view 

 

 

3 Wires 

4 Wires 

 

 

TEMPERATURE VSRESISTANCE TABLE (ITS90) 

°C 

Pt385 

Pt3926 

Pt3916 

-200°C 

18.521

 

16.996

 

17.057

 

-100°C 

60.256

 

59.479

 

59.565

 

0°C 

100.000

 

100.000

 

100.000

 

100°C 

138.505

 

139.272

 

139.171

 

200°C 

175.856

 

177.362

 

177.155

 

300°C 

212.052

 

214.275

 

213.957

 

400°C 

247.092

 

250.018

 

249.584

 

500°C 

280.977

 

284.591

 

284.036

 

600°C 

313.708

 

317.994

 

317.313

 

700°C 

345.280

 

800°C 

375.700

 

 

OPERATOR MAINTENANCE 

WARNING 

To avoid possible electrical shock, disconnect the 

thermocouple connectors from the thermometer before 

removing the cover. 

Battery Replacement 

1. Power is supplied by 4pcs 1.5V (AAA SIZE) UM-4 R03. 
2. The “

” appears on the LCD display when replacement 

is needed.  To  replace  battery  remove  screw  from back  of 
meter and lift off the battery cover. 

3. Remove the battery from battery contacts and replace. 
4. When not in use for long periods remove battery. 
5.  Do  not  store  in  locations  with  high  temperatures,  or  high 

humidity. 

Cleaning 

Periodically  wipe  the  case  with  a damp  cloth  and  detergent, 
do not use abrasives or solvents. 

*Software Operation manual is on the Software disk. 

 

WARRANTY / DISCLAIMER 

OMEGA  ENGINEERING,  INC.  warrants  this  unit  to  be  free  of  defects  in 
materials and workmanship for a period of 13 months from date of purchase. 
OMEGA’s WARRANTY adds an additional one (1) month grace period to the 
normal one (1) year product warranty to cover handling and shipping time. 
This ensures that OMEGA’s customers receive maximum coverage on each 
product. 

If  the  unit  malfunctions,  it  must  be  returned  to  the  factory  for  evaluation. 
OMEGA’s  Customer  Service  Department  will  issue  an  Authorized  Return 
(AR) number immediately upon phone or written request. Upon examination 
by OMEGA, if the unit is found to be defective, it will be repaired or replaced 
at  no  charge.  OMEGA’s  WARRANTY  does  not  apply  to  defects  resulting 
from  any  action  of  the  purchaser,  including  but  not  limited  to  mishandling, 
improper  interfacing, operation outside of design  limits,  improper  repair,  or 
unauthorized  modification.  This  WARRANTY  is  VOID  if  the  unit  shows 
evidence of having been tampered with or shows evidence of having been 
damaged  as  a  result  of  excessive  corrosion;  or  current,  heat,  moisture  or 
vibration;  improper  specification;  misapplication;  misuse  or  other  operating 
conditions  outside  of  OMEGA’s  control.  Components  which  wear  is  not 
warranted, include but are not limited to contact points, fuses, and triacs. 

OMEGA  is  pleased  to  offer  suggestions  on  the  use  of  its  various 
products.  However,  OMEGA  neither  assumes  responsibility  for  any 
omissions or errors nor assumes liability for  any damages that result 
from the use of its products in accordance with  information provided 
by  OMEGA,  either  verbal  or  written.  OMEGA  warrants  only  that  the 
parts  manufactured  by  it  will  be  as  specified  and  free  of  defects. 
OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF 
ANY  KIND  WHATSOEVER,  EXPRESS OR  IMPLIED,  EXCEPT THAT  OF 
TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY 
OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE 
ARE  HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies 
of  purchaser  set  forth  herein  are  exclusive,  and  the  total  liability  of 
OMEGA with respect to this order, whether based on contract, warranty, 
negligence,  indemnification,  strict  liability  or  otherwise,  shall  not 
exceed  the  purchase  price  of  the  component  upon  which  liability  is 
based. In no event shall OMEGA be liable for consequential, incidental 
or special damages.
 

CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor 
shall it be used: (1) as a “Basic Component” under 10 CFR 21 (NRC), used 
in or with any nuclear installation or activity; or (2) in medical applications or 
used  on  humans.  Should  any  Product(s)  be  used  in  or  with  any  nuclear 
installation or  activity,  medical application, used on  humans, or  misused  in 
any  way,  OMEGA  assumes  no  responsibility  as  set  forth  in  our  basic 
WARRANTY  /  DISCLAIMER  language,  and,  additionally,  purchaser  will 
indemnify OMEGA and hold OMEGA harmless from any liability or damage 
whatsoever arising out of the use of the Product(s) in such a manner. 
 

RETURN REQUESTS / INQUIRIES 

Direct all warranty and repair requests / inquiries to the OMEGA Customer 
Service  Department.  BEFORE  RETURNING  ANY  PRODUCT(S)  TO 
OMEGA,  PURCHASER  MUST  OBTAIN  AN  AUTHORIZED  RETURN  (AR) 
NUMBER  FROM  OMEGA’S  CUSTOMER  SERVICE  DEPARTMENT  (IN 
ORDER  TO  AVOID  PROCESSING  DELAYS).  The  assigned  AR  number 
should  then  be  marked  on  the  outside  of  the  return  package  and  on  any 
correspondence. 
The  purchaser  is  responsible  for  shipping  charges,  freight,  insurance  and 
proper packaging to prevent breakage in transit. 

FOR WARRANTY RETURNS, 
please have the following 
information available BEFORE 
contacting OMEGA: 
1. Purchase Order number under 

which the product was 
PURCHASED, 

2. Model and serial number of the 

product under warranty, and 

3. Repair instructions and/or specific 

problems relative to the product. 

FOR NON-WARRANTY 
REPAIRS, consult OMEGA for 
current repair charges. Have the 
following information available 
BEFORE contacting OMEGA: 
1. Purchase Order number to cover 

the COST of the repair, 

2. Model and serial number of the 

product, and 

3. Repair instructions and/or specific 

problems relative to the product. 

OMEGA’s policy is to make running changes, not model changes, whenever 
an  improvement  is  possible.  This  affords  our  customers  the  latest  in  tech-
nology and engineering. 
OMEGA is a registered trademark of OMEGA ENGINEERING, INC. 

©  Copyright  2010  OMEGA  ENGINEERING,  INC.  All  rights  reserved.  This 
document  may  not  be  copied,  photocopied,  reproduced,  translated,  or 
reduced to any electronic medium or machine-readable form, in whole or in 
part, without the prior written consent of OMEGA ENGINEERING, INC. 
 

Where Do I Find Everything I Need 

for Process Measurement 

and Control? 

OMEGA…Of Course! 

Shop online at omega.com 

 

TEMPERATURE 

  Thermocouple, RTD & Thermistor Probes, 

Connectors, Panels & Assemblies 

  Wire: Thermocouple, RTD & Thermistor 
  Calibrators & Ice Point References 
  Recorders,  Controllers & Process Monitors 
  Infrared Pyrometers 

 

PRESSURE, STRAIN AND FORCE 

  Transducers & Strain Gages 
  Load Cells & Pressure Gages 
  Displacement Transducers 
  Instrumentation & Accessories 

 

FLOW/LEVEL 

  Rotameters, Gas Mass Flowmeters & Flow 

Computers 

  Air Velocity Indicators 
  Turbine/Paddlewheel Systems 
  Totalizers & Batch Controllers 

 

pH/CONDUCTIVITY 

  pH Electrodes, Testers & Accessories 
  Benchtop/Laboratory Meters 
  Controllers, Calibrators, Simulators & Pumps 
  Industrial pH & Conductivity Equipment 

 

DATA ACQUISITION 

  Data Acquisition & Engineering Software 
  Communications-Based Acquisition Systems 
  Plug-in Cards for Apple, IBM & Compatibles 
  Datalogging Systems 
  Recorders, Printers & Plotters 

 

HEATERS 

  Heating Cable 
  Cartridge & Strip Heaters 
  Immersion & Band Heaters 
  Flexible Heaters 
  Laboratory Heaters 

 

ENVIRONMENTAL 
MONITORING AND CONTROL 

  Metering & Control Instrumentation 
  Refractometers 
  Pumps & Tubing 
  Air, Soil & Water Monitors 
  Industrial Water & Wastewater Treatment 
  pH, Conductivity & Dissolved Oxygen 

Instruments 

Made in Taiwan

 

Reviews: