background image

1-3

Addendum:  HHM29-IR

A

1.   Set the slide switch to TC.
2.   Plug in a K type surface thermocouple probe to the meter.
3.   Measure the target temperature using the surface probe.
4.   Unplug the surface probe from the meter
5.   Set the slide switch to IR.
6.   Open the laser door. Press and hold the momentary switch and

aim at the target.

7.   Adjust the Emissivity potentiometer until the meter reads the

same temperature measured via the thermocouple probe. Read
the new Emissivity setting.

TEMPERATURE MONITORING OVER TIME

The multi-meter can be set up for unattended temperature
monitoring. Open the battery door. There is a slide switch with a red
actuator on the right side of inside PC board. Set the switch to the
right. Now as long as the multi-meter is on and the selector switch is
set to °F/°C, and the outside slide switch is set to IR, the unit will
continue to measure temperature unattanded. The momentary
switch only turns on the laser in this mode. 

——————————            CAUTION   ——————————

• USE OF CONTROLS OR ADJUSTMENTS OR PERFORMANCE OF

PROCEDURES OTHER THAN THOSE SPECIFIED HERE MAY RESULT IN
HAZARDOUS RADIATION EXPOSURE.

• DO NOT LOOK AT THE LASER BEAM COMING OUT OF THE LENS OR

VIEW DIRECTLY WITH OPTICAL INSTRUMENTS. EYE DAMAGE CAN
RESULT.

• USE EXTREME CAUTION WHEN OPERATING THE LASER SIGHTING.
• NEVER POINT THE LASER BEAM AT A PERSON.
• KEEP OUT OF REACH OF ALL CHILDREN.
• DO NOT ATTEMPT TO OPEN THE MULTI-METER OR LASER SIGHTING.

THERE ARE NO USER-SERVICEABLE PARTS.

• DO NOT OPEN BATTERIES, DISPOSE OF IN FIRE, HEAT ABOVE 100 °C

(212 °F), EXPOSE CONTENTS TO WATER, RECHARGE, PUT IN
BACKWARDS, MIX WITH USED OR OTHER BATTERY TYPES. IT MAY
EXPLODE OR LEAK AND CAUSE PERSONAL INJURY.

• AFTER IR TEMPERATURE MEASUREMENT, SET THE SLIDE SWITCH

BACK TO “TC” TO SAVE BATTERY LIFE.

Summary of Contents for HHM29-IR Supermeter

Page 1: ... M U Totalizers Batch Controllers pH CONDUCTIVITY M U pH Electrodes Testers Accessories M U Benchtop Laboratory Meters M U Controllers Calibrators Simulators Pumps M U Industrial pH Conductivity Equipment DATA ACQUISITION M U Data Acquisition Engineering Software M U Communications Based Acquisition Systems M U Plug in Cards for Apple IBM Compatibles M U Datalogging Systems M U Recorders Printers ...

Page 2: ...OM THIS APERTURE LASER RADIATION AVOID DIRECT EYE EXPOSURE CLASS IIIa LASER PRODUCT COMPLIES WITH 21 CFR CHAPTER 1 SUBCHAPTER J DANGER MAX OUTPUT 5mW WAVE LENGTH 650 670 nm Danger and Certification Labels Laser Beam Aperture Laser Door Patent Pending Laser Momentary Switch Function Selector Switch Laser Power Indicator LED Slide Switch After IR temperature measurement set the slide switch back to ...

Page 3: ...es TEL 33 0 130 621 400 FAX 33 0 130 699 120 Toll Free in France 0800 4 06342 e mail france omega com Germany Austria Daimlerstrasse 26 D 75392 Deckenpfronn Germany TEL 49 0 7056 3017 FAX 49 0 7056 8540 Toll Free in Germany 0800 TC OMEGASM e mail germany omega com United Kingdom One Omega Drive Riverbend Technology Centre ISO 9002 Certified Northbank Irlam Manchester M44 5EX United Kingdom TEL 44 ...

Page 4: ... liable for consequential incidental or special damages CONDITIONS Equipment sold by OMEGA is not intended to be used nor shall it be used 1 as a Basic Component under 10 CFR 21 NRC used in or with any nuclear installation or activity or 2 in medical applications or used on humans Should any Product s be used in or with any nuclear installation or activity medical application used on humans or mis...

Page 5: ...ct temperature measurement 1 There is a slide switch on the right side of the multi meter Set the slide switch to TC 2 Plug a K type thermocouple probe SMP connection to the multi meter and measure the temperature Field of View Target ACCEPTABLE UNACCEPTABLE Figure 1 Field of View Positions Operational Flow Chart Long Term IR Measurement Open the Laser Door by Rotating Open the Battery Door Set th...

Page 6: ...ulti meter s field of view Using the laser sighting aim at the target and measure the temperature While measuring temperature via Infrared the thermocouple input becomes thermocouple output proportional to the temperature measured This thermocouple output K type can go to a strip chart recorder for temperature profiling See Figure 2 The line of sight of the multi meter does not coincide with that ...

Page 7: ...ector switch is set to F C and the outside slide switch is set to IR the unit will continue to measure temperature unattanded The momentary switch only turns on the laser in this mode CAUTION USE OF CONTROLS OR ADJUSTMENTS OR PERFORMANCE OF PROCEDURES OTHER THAN THOSE SPECIFIED HERE MAY RESULT IN HAZARDOUS RADIATION EXPOSURE DO NOT LOOK AT THE LASER BEAM COMING OUT OF THE LENS OR VIEW DIRECTLY WIT...

Page 8: ... 5 VDC Lithium batteries Battery Type Lithium Eveready Energizer model L91 Battery life Lithium 72 F 200 hours with no Laser sighting Ambient temperature 7 hours with Laser sighting Operating Temperature 0 to 40 C 32 to 104 F Operating Relative Humidity less than 70 without condensation Dimensions 7 9 x 3 5 x 2 3 200 x 90 x 59 mm Weight 1 3 lbs 600 g L A S E R S I G H T I N G Laser sighting A Lase...

Page 9: ...CLASS IIIa LASER PRODUCT COMPLIES WITH 21 CFR CHAPTER 1 SUBCHAPTER J DANGER MAX OUTPUT 5mW WAVE LENGTH 650 670 nm DO NOT STARE INTO BEAM OR VIEW DIRECTLY WITH OPTICAL INSTRUMENTS CLASS 3A LASER PRODUCT CONFORMS TO IEC825 1 1993 LASER RADIATION For additional specifications and general information regarding the multi meter refer to the User s Guide HHM29 Beam diameter 5 mm Beam Divergence 1mrad Pow...

Page 10: ... wrought iron 0 94 Iron and steel excluding stainless rusted iron plate 0 69 Iron and steel excluding stainless polished steel 0 07 Iron and steel excluding stainless polished steel oxidized at 1110 F 600 C 0 79 Iron and steel excluding stainless rolled sheet steel 0 66 Iron and steel excluding stainless rough steel plate 0 94 0 97 Lead gray and oxidized 0 28 Mercury 0 09 0 12 Molybdenum filament ...

Page 11: ...Marble polished light gray 0 93 Paints lacquers and varnishes Black matte shellac 0 91 Paints lacquers and varnishes aluminum paints 0 27 0 67 Paints lacquers and varnishes flat black lacquer 0 96 0 98 Paints lacquers and varnishes white enamel varnish 0 91 Porcelain glazed 0 92 Quartz opaque 0 68 0 92 Roofing Paper 0 91 Tape Masking 0 95 Water 0 95 0 96 Wood planed oak 0 90 NONMETALS 1 7 Addendum...

Reviews: