Wiring Options

Wiring Options

For 4-wire RTD probes, connect the four lead wires to your RTD logger as show in the figures below.

For 3-wire RTD probes, short inputs 3 and 4 together, then connect the lead wires to input 1, 2 and 3.
For 2-wire RTD probes, short inputs 3 and 4 together and inputs 1 and 2 together, then connect the RTD lead 
wires to inputs 2 and 3.

Warning: Note the polarity instructions. Do not attach wires to the wrong terminals. 100 

, 2 or 4 wire RTD probes 

are recommended for the most accurate performance. Most 100 

, 3 wire RTD probes will work, but Omega cannot 

guarantee the accuracy. To determine whether or not the 3-wire RTD probe will work, the resistance between the 
two same colored wires should be less than 1 

. (Note: Please contact the manufacturer of the RTD probe for 

questions on the resistance)

Installation Guide

Installing the Interface cable

- OM-CP-IFC200

  Insert the device into a USB port. The drivers will install automatically. 

-OM-CP-IFC110

  Plug the serial cable into the port and verify it is secure.

Installing the software

Insert the Omega Software Flash Drive in an open USB port. If the autorun does not appear, locate the drive 
on the computer and double click on 

Autorun.exe

. Follow the instructions provided in the Installation Wizard.

 

OM-CP-OCTRTD

KEY

4- Excitation Current Out (+)
3- Measurement (+) Input
2- Measurement(-) Input
1- Ground (-)

Summary of Contents for OM-CP-OCTRTD

Page 1: ...EET Xxxxx Xxxxxx Xxxxx Xxxxxx Xxxxxxxx M0000 0016 INSTRUCTION SHEET Shop online at omega comSM e mail info omega com Office Jl Radin Inten II No 62 Duren Sawit Jakarta 13440 Indonesia Workshop Jl Pahlawan Revolusi No 22B Jakarta 13430 Indonesia Phone 021 8690 6777 Hunting Fax 021 8690 6777 Mobile 62 816 1740 8925 Fax 021 8690 6771 ...

Page 2: ...uarantee the accuracy To determine whether or not the 3 wire RTD probe will work the resistance between the two same colored wires should be less than 1 Ω Note Please contact the manufacturer of the RTD probe for questions on the resistance Installation Guide Installing the Interface cable OM CP IFC200 Insert the device into a USB port The drivers will install automatically OM CP IFC110 Plug the s...

Page 3: ...is reached or the device is stopped At this point the device cannot be restarted until it has been re armed by the computer Downloading data from a data logger 1 Connect the logger to the interface cable 1 Highlight the data logger in the Connected Devices list Click Stop on the menu bar 2 Once the data logger is stopped with the logger highlighted click Download You will be prompted to name your ...

Page 4: ...ment 20 C to 60 C 4 F to 140 F 0 RH to 95 RH non condensing Material Anodized Aluminum Weight 15 9 oz 450 g Dimensions 3 5 in x 4 4 in x 1 5 in 89 mm x 112 mm x 39 mm OM CP OCTRTD General Specifications OM CP OCTRTD Battery Warning WARNING FIRE EXPLOSION AND SEVERE BURN HAZARD DO NOT SHORT CIRCUIT CHARGE FORCE OVER DISCHARGE CRUSH PENETRATE OR INCINERATE BATTERY MAY LEAK OR EXPLODE IF HEATED ABOVE...

Reviews: