OM-CP-QUADVOLT and OM-CP-OCTVOLT
OM-CP-OCTVOLT General Specifications
Description
OM-CP-OCTVOLT
Voltage Range
*See Table Below
Voltage Resolution
Voltage Accuracy
Memory 16,383/channel
Reading Rate 1 reading every second up to 1 reading every 12 hours
LED Indicator None
Channels 8
Required Interface Package IFC110 or IFC200
Baud Rate 2,400
Typical Battery Life 1 year
Operating Environment -40°C to +60°C, 0%RH to 95%RH (non–condensing)
Material Anodized aluminum
Dimensions 3.5 in x 4.4 in x 1.5 in
(89 mm x 112 mm x 39 mm)
Weight 17 oz (480 g)
Battery Warning
WARNING: FIRE, EXPLOSION, AND SEVERE BURN HAZARD. DO NOT SHORT CIRCUIT, CHARGE, FORCE OVER DISCHARGE, DISASSEMBLE, CRUSH,
PENETRATE OR INCINERATE. BATTERY MAY LEAK OR EXPLODE IF HEATED ABOVE 60°C (140°F).
*OM-CP-OCTVOLT Range, Resolution and Calibrated Accuracy
Nominal Range
+-100mV
0 to 2.5V
0 to 15V
0 to 30V
Measurement Range (VDC)
+-150mV
-0.25 to 2.75
-1 to 16
-2 to 32
Accuracy
+-0.01%FSR
+-0.01%FSR
+-0.10%FSR
+-0.10%FSR
Resolution (mV)
5µV
0.1
0.5
1.0
Common Mode Input Range
0 to 2.5V
0 to 2.5V
0 to 2.5V
0 to 2.5V