Troubleshooting Tips

Why is the wireless data logger not appearing in the software?

If the OM-CP-RFRHTemp2000A doesn’t appear in the Connected Devices panel, or an error message is  
received while using the OM-CP-RFRHTemp2000A, try the following:

 

• Check that the 

OM-CP-RFC1000-EXT is properly connected. For more information, see Troubleshooting 

Interface Cable problems 

(below)

.

• Ensure that the battery is not discharged. For best voltage accuracy, use a voltage meter connected to the 

battery of the device. If possible, try switching the battery with a new 9V lithium. 

• Ensure that no other 

Omega data logging software

 is running in the background.

• Ensure that the most current 

Omega data logging software

 is being used.

• Ensure that the 

Connected Devices 

panel is large enough to display devices. This can be verified by  

positioning the cursor on the edge of the Connected Devices panel until the resize cursor appears, then  
dragging the edge of the panel to resize it. The screen layout may also be reset in the options menu by  
selecting 

File

Options

, and scrolling to the bottom.

Troubleshooting Interface Cable problems 

Check that the software properly recognizes the connected OM-CP-RFC1000-EXT wireless receiver.

If the wireless data logger is not appearing in the 

Connected Devices

 list, it may be that the OM-CP-RFC1000 

-EXT is not properly connected.
1.  In the software, click the 

File

 button, then click 

Options

2.  In the 

Options

 window, click 

Communications

3.  The 

Detected Interfaces

 box will list all of the available communication interfaces. If the OM-CP-

RFC1000-EXT is listed here, then the software has correctly recognized and is ready to use it.

Check that Windows recognizes the connected OM-CP-RFC1000-EXT wireless receiver.

If the software does not recognise the OM-CP-RFC1000-EXT, there may be a problem with Windows or the 
USB drivers.
1.  In Windows, click 

Start

, right-click 

Computer

 and choose 

Properties

 or press 

Break

 as  

a keyboard shortcut. 

2.  Click 

Device Manager

 in the left hand column.

3.  Double click 

Universal Serial Bus Controllers

4.  Look for an entry for

 Data logger Interface

.

5.  If the entry is present, and there are no warning messages or icons, then windows has correctly  

recognized the connected OM-CP-RFC1000-EXT. 

6.  If the entry is not present, or has an exclamation point icon next to it, the USB drivers may need to be 

installed. These are available on the software flash drive included with the OM-CP-RFC1000-EXT, and on  
the Omega website.

Ensure that the USB end of the OM-CP-RFC1000-EXT is securely connected to the computer.

1.  Locate the USB-A plug of the OM-CP-RFC1000-EXT. 
2.  If the interface cable is connected to the PC, unplug it. Wait ten seconds. 
3.  Reconnect the cable to the PC.
4.  Check to make sure that the red LED is lit, indicating a successful connection.

OM-CP-RFC1000-EXT 

Summary of Contents for OM-CP-RFRHTemp2000A

Page 1: ...OM CP RFRHTEMP2000A Wireless Temperature and Humidity Data Logger...

Page 2: ...for 5 seconds to activate wireless communication LED Indicators Status Green LED blinks every 5 seconds to indicate unit is logging Wireless Blue LED blinks every 40 seconds to indicate unit is opera...

Page 3: ...nnected devices list showing that the device has been recognized Select the logger in the connected devices list and click the claim button wait for the device to finish updating Select the Realtime S...

Page 4: ...Email and text alarms can also be configured and sent from within the software 1 To create an alarm rule select Manage Rules in the device tab of the software 2 Select New and enter a Rule name Enter...

Page 5: ...software click the File Button then click Options 2 In the Options window click Communications 3 The Detected Interfaces box will list all of the available communication interfaces If your OM CP RFC10...

Page 6: ...artment 3 Install the new battery taking note of the polarity 4 Push the cover closed until it clicks Recalibration Standard recalibration for the OM CP RFRHTemp2000A is one point at 25 C for the inte...

Page 7: ...eceiver Sensitivity OM CP RFC1000 EXT 95dBm typical Wireless Range 2000 max outdoors line of sight unobstructed 500 max indoors typical urban Sample Rate 1 reading every second up to 1 reading every 2...

Page 8: ...OM CP RFC1000 EXT Wireless Transceiver...

Page 9: ...her Omega devices Installation Guide Installing the Software Insert the Software CD or USB Flash Drive into a Windows PC If the autorun does not appear locate the drive on the computer and double clic...

Page 10: ...the OM CP RFC1000 EXT and the Omega data loggers should be mounted in the same orientation This usually means that the external antenna should be pointing straight up The antenna can pivot to accommod...

Page 11: ...data logger Software Locate and select the OM CP RFRHTemp2000A in the Connect Devices panel 4 In the Device tab click the Properties icon The Properties screen will display information about the devic...

Page 12: ...OM CP RFC1000 EXT is not properly connected 1 In the software click the File button then click Options 2 In the Options window click Communications 3 The Detected Interfaces box will list all of the a...

Page 13: ...ovals FCC ID OA3MRF24J40MC IC 7693A 24J40MC Transmission Distance To other RFC1000 EXTs 4000 max outdoor line of sight unobstructed 1000 max indoors typical urban environment Transmission Distance To...

Page 14: ...les aux appareils radio exempts de licence L exploitation est autoris e aux deux conditions suivantes 1 l appareil ne doit pas produire de brouillage et 2 l utilisateur de l appareil doit accepter tou...

Page 15: ...l application used on humans or misused in any way OMEGA assumes no responsibility as set forth in our basic WARRANTY DISCLAIMER language and additionally purchaser will indemnify OMEGA and hold OMEGA...

Reviews: