OM-CP-RFRHTemp2000A
Product Maintenance
Battery Replacement
Materials: OM-CP-BAT103
1. On the bottom of the enclosure, open the battery compartment by pulling in on the cover tab.
2. Remove the battery by pulling it from the compartment.
3. Install the new battery, taking note of the polarity.
4. Push the cover closed until it clicks.
Recalibration
Standard recalibration for the OM-CP-RFRHTemp2000A is one point at 25°C for the internal temperature
channel, and two points at 25%RH and 75%RH for the humidity channel. Recalibration is recommended
annually for any Omega data logger; a reminder is automatically displayed in the software when the device
is due.