69
SMS Notification Format
When the SMS notification is sent the recipient receives the SMS in the following format. (The
logger comment is displayed in the first row).
S/N 515244, Test 810#2, Int. Temperature 15/1/10 17:25
– High Alarm, Value Detected: 24.24C
5.17.
Calibration
All Omega data loggers are shipped fully calibrated either with a calibration certificate or a
Declaration of Accuracy certificate. However, DataSuite does provide a simple and efficient
process for users wishing to calibrate their data loggers themselves. The process itself is
performed with the OM-DAQLINK unit connected via USB cable. As with any data logger
calibration a calibrator or other type of calibration instrument is necessary.
The Calibration options are accessible via the Logger context menu in the Map View (right-
clicking the Logger icon).
Before performing logger calibration, the logger must be in
Stop
mode. In addition, all of the
calibration options are accessible with a password to prevent non-authorized users of the
system tampering
with the loggers’ calibration settings.
Note:
The default password is 1234. The password can be changed in the password dialog box.
The Calibration option menu options are:
Calibration:
Opens the main Calibration dialog
Save Calibration:
Saves the logger’s current calibration settings
Load Calibration:
Loads a logger’s previously saved calibration settings.
Reset Calibration:
Resets the logger’s calibration settings so that the raw hardware data is
obtained, without any software calibration applied to these values.
Restore Factory Calibration Default:
Restores the logger’s calibration settings to the
factory calibration i.e. the calibration settings it received prior to shipment from Omega.
5.17.1.
Introduction to Calibration
Although the data loggers come fully calibrated, the software enables you to calibrate any of the
logger sensors, on any input. The calibration parameters are sent to the data logger and stored
in its memory.
DataSuite employs two different calibration methods:
Two-point calibration
and
Offset
calibration.
Each sensor can be calibrated using the Two-point calibration method, and then tweaked using
offset calibration, except for the Thermocouples. These sensors require offset calibration only.