OM-WEB Series User's Guide
Specifications
40
Current excitation outputs (OM-WEB-TEMP)
Table 21. Current excitation output specifications
Parameter
Conditions
Specification
Configuration
4 dedicated pairs:
±IA: CH0/CH1
±IB : CH2/CH3
±IC: CH4/CH5
±ID: CH6/CH7
Current excitation output ranges
Thermistor
10 µA typ
RTD
210 µA typ
Tolerance
±5% typ
Drift
200 ppm/°C
Line regulation
2.1 ppm/V max
Load regulation
0.3 ppm/V typ
Output compliance voltage
(relative to AGND pins 9, 19, 28, 38)
3.90 V max
–0.03 V min
Note 18:
The excitation output currents should always be used in the dedicated configuration specified
above.
Note 19:
The current excitation outputs are automatically configured based on the sensor (thermistor or
RTD) selected.
Environmental
Table 22. Environmental specifications
Parameter
Conditions
Operating temperature range
0 °C to 55 °C
Storage temperature range
–40 °C to 85 °C
Humidity
0% to 90% non-condensing
Mechanical
Table 23. Mechanical specifications
Parameter
Conditions
Dimensions (L × W × H)
127 × 88.9 × 35.56 mm (5.0 × 3.50 × 1.40 in.)
Screw terminal connector type and pinout
Table 24. Screw terminal connector specifications
Parameter
Conditions
Connector type
Screw terminal
Wire gauge range
16 AWG to 30 AWG