Using the Handheld Infrared Thermometer
2
2-8
Figure 2-10. Recorder Hookup
5.
After all the data has been taken, press the
key to unlock
the trigger.
• Moving Surface Scan
- Measures the Temperature of Points on a
Moving Surface:
1.
Mount the thermometer on a camera tripod and aim at a fixed
point on the moving surface.
2.
Pull the trigger and press the
key to lock the trigger.
3.
If necessary, adjust the emissivity. The thermometer is now set
up for measuring the temperature of a moving surface.
4.
To record the temperature profile of the moving surface,
connect the IR thermometer to a strip chart recorder. Refer to
Figure 2-10 for details.
5.
After all data is taken, press the
key to unlock the trigger.
• Fixed Point Monitoring Over Time
- Monitors the temperature
at a fixed point over time:
NOTE: It is recommended that you use the ac adapter for long
term measurement of temperature.
1.
Mount the thermometer on a camera tripod and aim at the
target.
2.
Connect the analog output of the thermometer to a strip chart
recorder as shown in Figure 2-10.
3.
Pull the trigger and press the
key to lock the trigger.
4.
If necessary, adjust the emissivity.
5.
The thermometer is now set up for unattended monitoring of
temperature over time. You can also download the temperature
to a Serial Printer or a PC for further analysis (Models OS521
through OS524 only). Refer to Sections 2.3.7 and 2.3.8.
6.
After all data is taken, press the
key to unlock the trigger.
To Strip Chart
Recorder
Analog
Cable
NOTE
Center hole is the
analog output jack
Summary of Contents for OMEGASCOPE OS520
Page 4: ...ii Unpacking Instructions This page is intentionally blank ...
Page 7: ...v TABLE OF CONTENTS ...
Page 13: ...General Description 1 1 6 Notes ...
Page 49: ...4 3 Sighting Scope 4 Notes ...
Page 57: ...Troubleshooting Guide 6 6 6 Notes ...
Page 63: ...Glossary of Key Strokes 8 8 2 Notes ...
Page 73: ...Appendix Determining an Unknown Emissivity C C 4 Notes ...
Page 77: ...I 4 Index I Notes ...