6-5
Troubleshooting Guide
6
The Laser Sight Module
circle is not the right
diameter.
Laser Sight Module “line
of sight” does not coincide
with the center of the
target.
1. You do not have the correct Laser
Sight Module to match the
particular model of OS520
Series thermometer. Check the
identification label on the Laser
Sight Module. Make sure the
model number matches the model
number of your thermometer.
If not, call our Customer Service
Department.
1. The line of sight and the center
of the target are offset by design.
(refer to Figure 3-6 and the
explanation above it for how
to compensate for this).
2. The Laser Sight Module is not
properly assembled to the
thermometer (refer to Section
3.3.1 for the proper installation
procedure).
LASER SIGHT MODULE
Problem
Solution
No Laser Beam
1. Make sure the power contacts on
the top of the thermometer are
clean.
2. Correctly install the Laser Sight
Module. Refer to Chapter 3.
3. Make sure the trigger is pulled and
the laser power switch is pressed.
(The red power LED should be lit).