11.. . A
PPENDIX
A: M
ASTER
I
CON
R
EFERENCE
U
SING THE
OMP-MODL
11-83
NOTE: The link (connection to the Alternate Rate Input on the Sample
Rate Clock is unique in that it feeds back from an Output terminal to an
Input terminal that is to the left on the screen. This is the only link that can
be connected this way.
To make the connection, drag the link from the
Sample Rate Clock Input terminal to the Output terminal (left to right).
Enable:
Processing of icon is allowed when Enable pin is unconnected or
when connected
and
Enable signal is TRUE.
The Enable terminal can be left unconnected, connected to a Start/Stop
Clock icon, or connected to any other logic signal. This could control the
generation of Update signals as a condition of some other parameter.
OUTPUTS:
Output Signal:
Update type. The Sample Rate Clock icon is the only icon that
can generate an Update command/signal. (The Warm-up icon does not
actually generate an Update signal, it just passes it through).
ICON CONFIGURATION DIALOG BOX:
CONFIGURATION OPTIONS:
Icon Name:
Specify the label for the icon. This name will show directly under the icon
within the Program Net.
Never use Alternate Rate / Use Alternate Rate when _____ is True:
If the Alternate
Rate Input is left unconnected,
Never use Alternate Rate
will be displayed.
When a link is made to the Alternate Rate Input,
Use Alternate Rate when
`_____’ is True
is displayed. In quotes, HyperNet displays the Output
Name of the icon connected to the Alternate Rate Input terminal. The
displayed Output Name can be changed in the other icon’s dialog.
Update pulse every:
A text box is provided for User entry of a number. The
units (seconds, minutes, hours, days) can be specified via the list box.
NOTE:
To specify Update signal rates faster than 1 second, set the
Sample Rate Clock Resolution
to
milliseconds
from within the Global
icon Configuration dialog. Use of the millisecond setting will result in
higher power consumption as the OMP-MODL microprocessor is
continually running during this mode. In the
Seconds
mode, the
microprocessor is put to a low power `sleep’ mode during times of
inactivity.
NOTE:
In MILLISECOND Mode, logging sessions must be
limited to a maximum length of 2 weeks. This limitation
Summary of Contents for OMP-MODL
Page 1: ......
Page 9: ...1 INTRODUCTION USING THE MODULOGGER 1 7 13 ...
Page 23: ...2 OMP MODL System Base Using the OMP MODL 2 14 NOTES ...
Page 24: ...2 ModuLogger System Base Using the ModuLogger 2 1 ...
Page 30: ...3 INTERFACE MODULES USING THE OMP MODL 3 6 NOTES ...
Page 70: ...3 INTERFACE MODULES USING THE OMP MODL 3 46 NOTES ...
Page 76: ...3 INTERFACE MODULES USING THE OMP MODL 3 52 NOTES ...
Page 82: ...3 INTERFACE MODULES USING THE OMP MODL 3 58 NOTES ...
Page 83: ...3 INTERFACE MODULES USING THE MODULOGGER 3 1 ...
Page 105: ...5 HYPERCOMM COMMUNICATIONS USING THE OMP MODL 5 18 NOTES ...
Page 106: ...5 HYPERCOMM COMMUNICATIONS USING THE MODULOGGER 5 1 ...
Page 117: ...6 PCMCIA CARD CONFIGURATION AND USE USING THE MODULOGGER 6 1 ...
Page 134: ...7 HYPERNET ICON BASED PROGRAMMING USING THE OMP MODL 7 17 NOTES ...
Page 141: ...8 P P C D U OMP MODL ...
Page 149: ...9 HYPERPLOT GRAPHIC DATA DISPLAY USING THE OMP MODL 9 8 ...
Page 161: ...10 HYPERTRACK REAL TIME DATA DISPLAY USING THE OMP MODL 10 12 NOTES ...
Page 162: ...10 HYPERTRACK REAL TIME DATA DISPLAY USING THE OMP MODL 10 1 ...
Page 163: ...11 APPENDICES USING THE OMP MODL 11 1 11 APPENDICES ...
Page 164: ...11 APPENDICES USING THE OMP MODL 11 2 NOTES ...
Page 270: ...11 APPENDIX A MASTER ICON FILE REFERENCE USING THE OMP MODL 11 108 ...
Page 276: ...11 APPENDIX C HYPERWARE FILE LISTING USING THE OMP MODL 11 114 NOTES ...
Page 278: ...11 APPENDIX C HYPERWARE FILE LISTING USING THE OMP MODL 11 116 NOTES ...
Page 279: ......
Page 282: ...11 APPENDIX E CHANGING THE OMP MODL EPROM USING THE OMP MODL 11 120 NOTES ...
Page 288: ...11 APPENDIX G HYPERNET THEORY OF OPERATION USING THE OMP MODL 11 126 NOTES ...
Page 290: ...11 APPENDIX I RS 232 CABLE PORT AND ADAPTER USING THE OMP MODL 11 128 NOTES ...
Page 297: ...11 APPENDIX K MODEM CONFIGURATION USING THE OMP MODL 11 135 ...