Precautions before first use 

1. Install batteries before use
2. Before first use, warm up the device for about 10 minutes before installing the battery.
3. When using it for body temperature measurement point thermometer at the forehead.  This is a

non-contact thermometer and the suggested measuring distance to the forehead is 2” (5 cm).
Press the trigger once to take one temperature measurement.  Tip: temperature can also be
measured at an earlobe.

4. When measuring body temperature, please make sure there are no obstacles between the IR

sensor and forehead skin, e.g., move hair or sweat if obstructing measurement.

Warnings 

The glass protector in front of the IR sensor is a critical but very fragile element of the device.

Use cotton and 70% rubbing alcohol for cleaning.

Please only use qualified battery. Do not charge batteries that are not rechargeable.

Do not use the thermometer under the sun or under water.

Maintenance 

Please follow the instruction manual for troubleshooting. If the issue is not listed in the
manual, please contact our customer support.

Warning sign “HI”

When using the thermometer for measuring body temperature, the “HI” warning sign may
come on. The warning sign indicates the temperature measured exceeds the maximum
allowed threshold 108.5⁰F (42.5⁰C).

Warning sign “LO”

When using the thermometer for measuring body temperature, the “LO” warning sign may
come one. The warning sign indicates the temperature measured exceeds the minimum
allowed threshold 89.6⁰F (32.0⁰C).

Troubleshooting 

When using the thermometer in body temperature mode, if a “HI” or “Lo” warning sign is
displayed and it is suspected that the temperature is incorrect please follow these steps:

1.)

 

Switch to surface temperature mode, measure temperature of any surface. 

2.)

 

If the temperature looks correct, then go to menu (F2). 

3.)

 

Check the offset, the offset should be set to 0⁰C (0⁰F). 

4.)

 

If offset not set to 0⁰C (0⁰F), then set offset to 0⁰C (0⁰F). 

5.)

 

Switch back to body temperature mode and measure body temperature again, body 
temperature should now be accurate. 

Summary of Contents for OS-821 Series

Page 1: ...User s Guide Shoponlineat omega com e mail info omega com For latest product manuals omega com en us pdf manuals OS820 Series Non Contact Body IR Thermometer with 3 Color Alarm Backlight Display TM ...

Page 2: ...otice Servicing North America U S A Omega Engineering Inc Headquarters 800 Connecticut Ave Suite 5N01 Norwalk CT 06854 omega com info omega com Toll Free 1 800 826 6342 USA Canada only Customer Service 1 800 622 2378 USA Canada only Engineering Service 1 800 872 9436 USA Canada only Tel 203 359 1660 Fax 203 359 7700 e mail info omega com For Other Locations Visit omega com worldwide ...

Page 3: ......

Page 4: ... indicate measured temperature 2 Backlight button turn on off backlight Hold to change 3 Left Button during the measurement mode page up the stored temperature data records activate the parameter setting mode or during the parameter setting mode navigate the settings menu 4 Right Button during the measurement mode page down the stored temperature data records or during the parameter setting mode a...

Page 5: ...erature Units from C to F F1 Press and hold SET left button until you see F1 appear in bottom right corner of screen While showing F1 Symbol in bottom right corner of the display press ADJ to switch between C and F Temperature Offset Setting F2 With F2 Symbol showing in bottom right corner of the display press ADJ to increase and MODE to decrease to adjust the temperature offset within the range o...

Page 6: ...o take one measurement The number on the button left corner shows the record number The backlight color indicates the different temperature range as follows Green Below 99 5 F 37 5 C Orange 99 5 100 4 F 37 5 C 38 C Red Above 100 4 F 38 C Data Record Management View Data Record Press SET left button and use ADJ right button to page up and down among the data records Delete Data Record Press and hol...

Page 7: ...tenance Please follow the instruction manual for troubleshooting If the issue is not listed in the manual please contact our customer support Warning sign HI When using the thermometer for measuring body temperature the HI warning sign may come on The warning sign indicates the temperature measured exceeds the maximum allowed threshold 108 5 F 42 5 C Warning sign LO When using the thermometer for ...

Page 8: ...GA if the unit is found to be defective it will be repaired or replaced at no charge OMEGA s WARRANTY does not apply to defects resulting from any action of the purchaser including but not limited to mishandling improper interfacing operation outside of design limits improper repair or unauthorized modification This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows ...

Page 9: ... Turbine Paddlewheel Systems M U Totalizers Batch Controllers pH CONDUCTIVITY M U pH Electrodes Testers Accessories M U Benchtop Laboratory Meters M U Controllers Calibrators Simulators Pumps M U Industrial pH Conductivity Equipment DATA ACQUISITION M U Communications Based Acquisition Systems M U Data Logging Systems M U Wireless Sensors Transmitters Receivers M U Signal Conditioners M U Data Acq...

Reviews: